Blue Elephant Inc
6716 Greenleaf Ave, Whittier, California, United States
FEI 3031025053
FDA has one inspection on file for Blue Elephant Inc in Whittier, California (FEI 3031025053). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 9 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2025
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Nov 17, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training records
Personnel training is not documented.
Design input - documentation
Design input requirements were not adequately documented.
Acceptable supplier records, inadequate records
Records of acceptable suppliers and consultants have not been adequately established.
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been established.
Design review - documentation
The design review results, including identification of the design, the date and the individual(s) performing the review, were not documented in the design history file.
Show 4 more citations
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Design verification - documentation
The design verification results, including identification of the design, method(s), the date and the individual(s) performing the verification, were not documented in the design history file.
DMR - not or inadequately maintained
A device master record has not been maintained.
Design output - documentation
Design output was not documented before release.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Nov 17, 2025. Each one asked for a Form 483.
Inspection ending May 16, 2025VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 8, 2025. FDAgent holds no copy that FDA posted.
- Nov 17, 2025FDA-FOIA-2025-13040Form 483· 1 of 4 sitesRecords released
- Jul 16, 2025FDA-FOIA-2025-8757Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 4 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.25Personnel1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.181Device master record1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3031025053 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3031025053. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.