Skip to content

Triangle Machine, Inc.

3702 W Sample St Ste 1125, South Bend, Indiana, United States

FEI 3015526568

FDA has 2 inspections on file for Triangle Machine, Inc. in South Bend, Indiana (FEI 3015526568), from 2022 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 7 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
2, 2022–2026
0 OAI · 2 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Jun 23, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.3.8

      Design and development transfer procedures not documented

      Procedures for transfer of design and development outputs to manufacturing were not adequately documented.

    2. ISO 13485:2016 Clause 7.5.1

      Production controls not adequate

      Production controls do not include documentation of procedures and methods for the control of production.

    3. ISO 13485:2016 Clause 8.1

      Methods for monitoring, measurement, analysis, and improvement not determined

      Appropriate methods, including statistical techniques, and the extent of their use were not adequately determined for monitoring, measurement, analysis, and improvement processes.

    4. ISO 13485:2016 Clause 4.1.4

      Changes to the QMS not adequately controlled or evaluated

      Changes to quality management system processes were not adequately controlled in accordance with requirements and evaluated for their impact on the quality management system and the medical devices produced.

    5. ISO 13485:2016 Clause 8.2.4

      Internal audit criteria, scope, interval, and methods not defined and recorded

      The internal audit criteria, scope, interval, and methods were not adequately defined and recorded.

    Show 2 more citations
    1. ISO 13485:2016 Clause 7.1

      Product realization risk management processes not documented

      One or more processes for risk management in product realization have not been documented.

    2. ISO 13485:2016 Clause 7.4.1

      Records of supplier (re)evaluation, selection, and monitoring not maintained

      Records of the results of evaluation, selection, monitoring and re-evaluation of supplier capability or performance were not adequately maintained.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    3. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    4. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    5. Management review - Lack of or inadequate procedures

      Procedures for management review have not been established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Jun 23, 2026. It asked for a Form 483.

  • Inspection ending Apr 15, 2026VAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 21, 2026. FDAgent holds no copy that FDA posted.

    1. Jun 23, 2026FDA-FOIA-2026-6011Form 483· 1 of 4 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 5 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3015526568 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3015526568. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.