Alaris Biologic Technologies, Inc.
3542 Nw 97th Blvd, Gainesville, Florida, United States
FEI 3010898668
FDA has 3 inspections on file for Alaris Biologic Technologies, Inc. in Gainesville, Florida (FEI 3010898668), from 2015 to 2018. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 3 citations from this inspection, all under 21 CFR Part 1271, the rules for human cell and tissue products. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 1271.230(c) and 21 CFR 1271.160(c).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2015–2018
- 0 OAI · 3 VAI · 0 NAI
- Inspections with 483 observations shown
- 3 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documented, dated, signed
The validation activities were not documented.
Changes to validated process
A validated process that was changed was not reviewed and evaluated.
Deviations not reported to FDA
HCT/P deviations relating to core CGTP requirements that occurred in your establishment were not reported to FDA.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality audits performed periodically
Periodic quality audits of activities related to core CGTP requirements have not been performed.
All core requirements covered in program
A quality program which addresses all of the core CGTP requirements, appropriate for the HCT/Ps manufactured and the manufacturing steps performed, has not been maintained , documented , followed and revised.
Documentation of maintenance and cleaning
Documentation of equipment maintenance, cleaning, sanitization, and calibration was not maintained.
Back-up of electronic records
Records stored on electronic systems are not backed up.
Ensurane of compliance
You did not ensure that establishment(s) that by contract, agreement or arrangement, perform manufacturing steps for you were in compliance with applicable CGTP requirements prior to the initiation of the contract, agreement of arrangement.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation not maintained
Documentation of environmental control and monitoring activities was not maintained.
Adequate control, proper conditions
Environmental conditions existed in which contamination or cross contamination of HCT/Ps or equipment could occur, and proper conditions for operations were not provided.
Inspected routinely
Equipment was not routinely inspected for adherence to maintenance schedules.
Release criteria verified, documented
Release criteria were not documented to have been met through a review of manufacturing and tracking records before HCT/Ps were made available for distribution.
Responsibility for review before distribution
HCT/Ps were determined to meet all release criteria and made available for distribution without review of manufacturing records to determine that all applicable requirements were met.
Show 7 more citations
Records incomplete
Records did not show the dates of entries and were not detailed as necessary to provide a complete history of work performed.
Not qualified, authorized for functions
Personnel perform functions for which they are not qualified and authorized.
Documentation of activities
Documentation of facility cleaning and sanitation activities was not maintained.
Changes to validated process
A validated process that was changed was not reviewed and evaluated.
Clean, sanitary, orderly manner
Facilities were not maintained in a clean manner to prevent the introduction, transmission, or spread of communicable disease.
Description of methods, schedule
Facility cleaning and sanitation procedures did not describe the cleaning methods to be used and the schedule for cleaning.
Quality audits performed periodically
Periodic quality audits of activities related to core CGTP requirements have not been performed.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 1271.150Current good tissue practice requirements2 of 3
- 21 CFR 1271.160Establishment and maintenance of a quality program2 of 3
- 21 CFR 1271.200Equipment2 of 3
- 21 CFR 1271.230Process validation2 of 3
- 21 CFR 1271.270Records2 of 3
- 21 CFR 1271.170Personnel1 of 3
- 21 CFR 1271.190Facilities1 of 3
- 21 CFR 1271.195Environmental control and monitoring1 of 3
Show all 10
- 21 CFR 1271.265Receipt, predistribution shipment, and distribution of an HCT/P1 of 3
- 21 CFR 1271.350Reporting1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3010898668 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3010898668. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.