FDA has one inspection on file for Healgen Scientific Llc in Houston, Texas (FEI 3009238284). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
5 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2016
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 5 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures for storage.
Procedures for the control of storage areas and stock rooms have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been adequately established.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Regulations FDA cited here
The 4 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 1
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 1
- 21 CFR 820.150Storage1 of 1
- 21 CFR 820.198Complaint files1 of 1
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3009238284, one row per refused product line.
5 refusals from 2021-02-26 to 2022-04-08 · charged NO 510(K) on 5, NO PMA on 2 · no FDA sample analysis
FDA firm profile- 2022-04-08
COVID-19 MULTI-ANALYTE ANTIGEN DEVICE83QMN
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports799-7220259-1/21/1
- 2022-02-22
COVID-19 TEST HOME COLLECTION KIT DEVICES83QLW
NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports799-7212313-6/11/1
- 2022-02-15
REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID83QJR
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports799-7266441-0/11/1
- 2021-12-16
REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID83QJR
NOT LISTED 502(o), 801(a)(3); MISBRANDING
NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports799-6431619-3/11/1
- 2021-02-26
CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM.83QKP
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Northeast ImportsAEK-8743874-0/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3009238284 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3009238284. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.