FDA has one inspection on file for Jjgc Industria E Comercio De Materials Dentarios Sa in Curitiba, Brazil (FEI 3008261720). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 4 citations from this inspection: 2 under 21 CFR Part 803, the device adverse event reporting rules, and 2 under 21 CFR Part 820, the device quality system rules. They are listed below.
2 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2017
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 2 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Records of complaint investigation
Records of complaint investigations do not include required information.
Software validation for automated processes
Software used as part of the quality system has not been validated for its intended use according to an established protocol.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Conducting an Investigation
An MDR report with incomplete information was submitted to FDA without a statement explaining why such information was incomplete and the steps taken to obtain the information.
Regulations FDA cited here
The 3 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 1
- 21 CFR 820.70Production and process controls1 of 1
- 21 CFR 820.198Complaint files1 of 1
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3008261720, one row per refused product line.
2 refusals from 2024-02-13 to 2025-07-11 · charged NOT LISTED on 2, FRNMFGREG on 2 · no FDA sample analysis
FDA firm profile- 2025-07-11
IMPLANT, ENDOSSEOUS76DZE
NOT LISTED 502(o), 801(a)(3); MISBRANDING
RXPERSONAL 502(a), 502(f)(1), 801(a)(3), MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsAEK-8546075-4/12/1
- 2024-02-13
IMPLANT, ENDOSSEOUS76DZE
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsAEK-4551926-9/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3008261720 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3008261720. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.