Skip to content

Hanuri Distribution Inc

9601 Owensmouth Ave Ste 8, Chatsworth, California, United States

FEI 3007778244

FDA has 4 inspections on file for Hanuri Distribution Inc in Chatsworth, California (FEI 3007778244), from 2013 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 4 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.80(d).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2013–2026
0 OAI · 3 VAI · 1 NAI
Inspections with 483 observations shown
3 of 4
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Jun 16, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.4.1

      Procedures for purchasing process not documented

      Procedures to ensure that purchased product conforms to specified purchasing information have not been documented.

    2. ISO 13485:2016 Clause 8.3.1

      Procedure for nonconforming product not documented

      Procedures to define the controls and related responsibilities and authorities for the identification, documentation, segregation, evaluation, and disposition of nonconforming product were not documented.

    3. ISO 13485:2016 Clause 5.6.1

      Management review procedures not documented

      Procedures for management review have not been documented.

    4. ISO 13485:2016 Clause 7.3.5

      Records of design and development reviews not maintained

      Records of the results of the design and development reviews and any necessary actions were not maintained or include identification of the design under review, the participants involved, and the date of the review.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Equipment control activity documentation

      Equipment calibrations have not been documented.

    2. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

    3. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    4. Lack of procedures, or not maintained

      Document control procedures have not been established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Jun 16, 2026. Each one asked for a Form 483.

  • Inspection ending Mar 26, 2026VAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 10, 2026. FDAgent holds no copy that FDA posted.

    1. Jun 16, 2026FDA-FOIA-2026-5853Form 483· 1 of 4 sitesRecords released
    2. May 6, 2026FDA-FOIA-2026-4586Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3007778244 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3007778244. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.