Skip to content

Kyungwon Medical Co., Ltd.

765, Jeongjungyeonje-ro, Osong-eup, Korea (the Republic of)

FEI 3006455376

FDA has one inspection on file for Kyungwon Medical Co., Ltd. in Osong-eup, Korea (the Republic of) (FEI 3006455376). It ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 7 citations from this inspection: 6 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.

FDA’s Red Lists for Import Alerts 89-04 and 89-08 name this firm at this address. FDA printed no FEI on the entries. No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
1, in 2012
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
89-04, 89-08Red List
Matched on FDA’s firm name and street address

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record.

    2. Design validation - production units

      The design was not validated using initial production units, lots or batches or their equivalents.

    3. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    4. Equipment Installation, Placement, Specified Requirements

      The appropriate design, construction, placement, and installation of manufacturing equipment have not been ensured.

    5. Process changes - review, evaluation and revalidation

      A validated process was not revalidated when changes or process deviations occurred.

    Show 2 more citations
    1. Design output - documentation

      Design output was not adequately documented before release.

    2. Document change records, maintained.

      Records of changes to documents were not adequately maintained.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Import alerts

FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.

  • 89-08Red ListDetention Without Physical Examination of Devices without Approved PMA's or IDE's and Other Devices Not Substantially Equivalent or Without a 510(k)Import alert pageAlert on fda.gov

    Date published 2016-01-27

    FDA’s entry
    Kyungwon Medical Co., Ltd. · 765, Jeongjungyeonje-ro, Osong-eup, Chungcheongbuk-do KOREA (THE REPUBLIC OF)
    FDA’s inspection record for FEI 3006455376
    Kyungwon Medical Co., Ltd. · 765, Jeongjungyeonje-ro, Osong-eup, Korea (the Republic of)

    Matched on exact firm name, street address and country. FDA printed no FEI on this entry.

    2 products

    • 87 M - - BPFiller, Bone Void, Osteoinduction (w/o Human Growth Factor)
    • 87 M - - QVFiller, Calcium Sulfate Preformed Pellets
  • 89-04Red ListDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsImport alert pageAlert on fda.gov

    Date published 2013-01-07

    FDA’s entry
    Kyungwon Medical Co., Ltd. · 765, Jeongjungyeonje-ro, Osong-eup, Chungcheongbuk-do KOREA (THE REPUBLIC OF)
    FDA’s inspection record for FEI 3006455376
    Kyungwon Medical Co., Ltd. · 765, Jeongjungyeonje-ro, Osong-eup, Korea (the Republic of)

    Matched on exact firm name, street address and country. FDA printed no FEI on this entry.

    2 products

    • 84 G - - XPMethyl Metacrylate For Cranioplasty
    • 87 M - - NHOrthosis, Spondyloisthesis Spinal Fixation

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3006455376 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3006455376. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.