Sun Dental Laboratories, LLC
1800 9th Ave N, Saint Petersburg, Florida, United States
FEI 3006430881
FDA has 2 inspections on file for Sun Dental Laboratories, LLC in Saint Petersburg, Florida (FEI 3006430881), from 2023 to 2026. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 8 citations from this inspection. They are listed below. Four of the six FOIA requests for this site have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 803.17.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 2, 2023–2026
- 1 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 6 requests, latest Sep 15, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 4.1.3
QMS processes not adequate
For each quality management system process, your firm did not determine criteria and methods needed to ensure that both the operation and control of these processes are effective, ensure the availability of resources and information necessary to support the operation and monitoring of these processes, implement actions necessary to achieve planned results and maintain the effectiveness of these processes and monitor, measure, and analyze these processes.
- ISO 13485:2016 Clause 8.5.3
Preventive action procedures not documented
Preventive action procedures describing the requirements for determining potential nonconformities and their causes, evaluating the need for action to prevent occurrence of potential nonconformities, planning and documenting action needed and implementing such action, verifying that the action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of the preventive action taken have not been documented.
- ISO 13485:2016 Clause 8.5.2
Corrective action procedures not documented
Corrective action procedures defining the requirements for reviewing nonconformities, determining the causes of nonconformities, evaluating the need for action to ensure that nonconformities do not recur, planning and documenting action needed and implementing such action, verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of corrective action taken have not been documented.
UDI was not recorded for each medical device or batch
Unique device identification (UDI) was not recorded for each medical device or batch of medical devices.
- ISO 13485:2016 Clause 7.3.1
Design and development procedures not documented
Design and development procedures were not documented.
Show 3 more citations
- ISO 13485:2016 Clause 5.1
Lack of evidence for top management's commitment to QMS
Top management did not provide evidence of its commitment to the development and implementation of the quality management system and maintenance of its effectiveness by communicating to the organization the importance of meeting requirements, establishing the quality policy, ensuring the quality objectives are established, conducting management reviews and ensuring the availability of resources.
- ISO 13485:2016 Clause 7.1
Required elements not determined in planning product realization
In product realization planning your firm did not determine, as appropriate, quality objectives and requirements for the product, the need to establish processes and documents and to provide resources specific to the product, including infrastructure and work environment, required activities specific to the product together with criteria for product acceptance and records needed to provide evidence that the realization processes and resulting product meet requirements.
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained and implemented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Supplier notification of changes
There is no agreement with suppliers and contractors to notify you of changes in the product or service.
Labeling inspection - UDI
Labeling was not sufficiently examined by a designated individual for accuracy including unique device identifier (UDI) or universal product code (UPC).
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Six requests to FDA under the Freedom of Information Act named this site, most recently on Sep 15, 2026. Each one asked for a Form 483.
Inspection ending Apr 1, 2026OAI
4 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 4, 2026. FDAgent holds no copy that FDA posted.
- Sep 15, 2026FDA-FOIA-2026-8918Form 483· 1 of 5 sitesRecords released
- Sep 10, 2026FDA-FOIA-2026-8786Form 483Records released
- Jul 7, 2026FDA-FOIA-2026-6458Form 483· 1 of 4 sitesRecords released
- Jun 12, 2026FDA-FOIA-2026-5772Form 483· 1 of 5 sitesRecords released
Inspection ending Sep 20, 2023VAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 31, 2023FDA-FOIA-2023-9542Form 483Merged with another request
- Oct 30, 2023FDA-FOIA-2023-9537Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 4 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 2
- 21 CFR 820.35Control of records1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.120Device labeling1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3006430881 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3006430881. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.