BERLIN HEART GMBH
Wiesenweg 10, Berlin, Germany
FEI 3004582654
FDA has 2 inspections on file for BERLIN HEART GMBH in Berlin, Germany (FEI 3004582654), from 2013 to 2022. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 6 citations from this inspection: 4 under 21 CFR Part 820, the device quality system rules, 1 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 806, the device correction and removal reporting rules. They are listed below. One of the four FOIA requests for this site has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 2, 2013–2022
- 1 OAI · 0 VAI · 1 NAI
- Inspections with 483 observations shown
- 1 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 4 requests, latest Mar 17, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Project Evaluation: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Design input - Lack of or inadequate procedures
Procedures for design input have not been adequately established.
Design output - Lack of or inadequate procedures
Procedures for design output have not been adequately established.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Lack of System for Determining MDR Events
The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
Show 1 more citation
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Mar 17, 2025.
Inspection ending Oct 13, 2022OAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Oct 8, 2025. FDAgent holds no copy that FDA posted.
- Mar 17, 2025FDA-FOIA-2025-2404Form 483Records released
Inspection ending Jul 21, 2017
1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- Mar 4, 2022FDA-FOIA-2022-1790Inspection recordsClosed
Inspection ending Oct 17, 2014
1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- Jul 27, 2021FDA-FOIA-2021-4871Form 483Closed over fees
Inspection date not given
1 request. None of FDA’s log entries for these requests say records were released.
- Oct 27, 2020FDA-FOIA-2020-7666Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 806.10Reports of corrections and removals1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.184Device history record1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004582654 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004582654. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.