Bioland Technology Ltd.
No. A6B7(Block G) ShangRong Industrial Zone No.5 Baolong Road, Baolong Community Longgang District, Shenzhen, China
FEI 3004101980
FDA has one inspection on file for Bioland Technology Ltd. in Shenzhen, China (FEI 3004101980). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 6 citations from this inspection: 5 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
7 import refusals are filed under this FEI, the latest on . FDA’s Import Alert 89-08 Red List names this firm at this address. FDA printed no FEI on the entry. No court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 1, in 2014
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- 7 refusals
- Import alerts
- 89-08Red List
- Matched on FDA’s firm name and street address
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.
Lack of System for Determining MDR Events
The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Show 1 more citation
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 820.70Production and process controls1 of 1
- 21 CFR 820.75Process validation1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.184Device history record1 of 1
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3004101980, one row per refused product line.
7 refusals from 2023-04-19 to 2026-10-16 · charged NO 510(K) on 7, NOT LISTED on 5 · no FDA sample analysis
FDA firm profile- 2026-10-16
SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER75NBW
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southwest ImportsKDI-0066035-8/10002/1
- 2026-05-21
SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER75NBW
NOT LISTED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of West Coast ImportsNUP-1836436-7/33/1
- 2025-07-21
SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER75NBW
NOT LISTED 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsSCS-1071969-3/23/1
- 2025-07-21
SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER75NBW
NOT LISTED 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsSCS-1071969-3/13/1
- 2025-05-30
SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER75NBW
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsSCS-0125881-8/11/1
Show 2 more
- 2024-05-22
STRIP, DIALYSATE, PH, BICARBONATE, GLUCOSE, ACID, INDICATOR78NTZ
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsAEK-5060627-4/12/1
- 2023-04-19
SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER75NBW
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsAEK-3139517-9/11/1
Import alerts
FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.
- 89-08Red ListDetention Without Physical Examination of Devices without Approved PMA's or IDE's and Other Devices Not Substantially Equivalent or Without a 510(k)Import alert pageAlert on fda.gov
Date published 2025-02-06
- FDA’s entry
- Bioland Technology Ltd. · No. A6B7(Block G) ShangRong Industrial Zone No.5 Baolong Road, Baolong Community Longgang District, Shenzhen, Guangdong CHINA
- FDA’s inspection record for FEI 3004101980
- Bioland Technology Ltd. · No. A6B7(Block G) ShangRong Industrial Zone No.5 Baolong Road, Baolong Community Longgang District, Shenzhen, China
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
2 products
- 75 N - - BWSystem, Test, Blood Glucose, Over the Counter
- 75 N - - BWSystem, Test, Blood Glucose, Over the Counter
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004101980 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004101980. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.