AOMORI OLYMPUS CO., LTD. Kuroishi Site
2 Chome 248-1, Okkonoki, Kuroishi, Japan
FEI 3003995201
FDA has 3 inspections on file for AOMORI OLYMPUS CO., LTD. Kuroishi Site in Kuroishi, Japan (FEI 3003995201), from 2016 to 2025. The latest ended on and was classified NAI, no action indicated. FDA issued a warning letter on (CDRH).
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two of the four FOIA requests for this site have asked FDA for this inspection's 483, and one more for the records from a later inspection, ending Nov 12, 2025, that FDA's Data Dashboard doesn't list yet.
FDA’s Import Alert 89-04 Red List names this firm at this address. FDA printed no FEI on the entry. No court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.100(a) and 21 CFR 820.75(a).
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 3, 2016–2025
- 0 OAI · 1 VAI · 2 NAI
- Inspections with 483 observations shown
- 2 of 3
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 4 requests, latest Sep 25, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- 89-04Red List
- Matched on FDA’s firm name and street address
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Manufacturer Name and Address
An individual medical device manufacturer report submitted per Form FDA 3500A format did not contain in Block D the device manufacturer's name and address.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Sep 25, 2026.
Inspection ending Nov 12, 2025
1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- May 1, 2026FDA-FOIA-2026-4360EIR· 1 of 8 sitesClosed
Inspection ending Nov 7, 2025NAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jun 18, 2026. FDAgent holds no copy that FDA posted.
- Sep 25, 2026FDA-FOIA-2026-9319Form 483 + EIROpen
- Jan 14, 2026FDA-FOIA-2026-497Form 483· 1 of 2 sitesRecords released
Inspection ending Jul 14, 2022VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Feb 5, 2024FDA-FOIA-2024-1255Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.75Process validation2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 803.52If I am a manufacturer, what information must I submit in my individual adverse event reports?1 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.184Device history record1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Import alerts
FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.
- 89-04Red ListDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsImport alert pageAlert on fda.gov
Date published 2026-10-19
- FDA’s entry
- AOMORI OLYMPUS CO., LTD. Kuroishi Site · 2 Chome 248-1, Okkonoki, Kuroishi, Aomori JAPAN
- FDA’s inspection record for FEI 3003995201
- AOMORI OLYMPUS CO., LTD. Kuroishi Site · 2 Chome 248-1, Okkonoki, Kuroishi, Japan
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
2 products
- 77 E - - OQBronchoscope (Flexible Or Rigid)
- 78 F - - GBUreteroscope
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003995201 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003995201. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.