Skip to content

AOMORI OLYMPUS CO., LTD. Kuroishi Site

2 Chome 248-1, Okkonoki, Kuroishi, Japan

FEI 3003995201

FDA has 3 inspections on file for AOMORI OLYMPUS CO., LTD. Kuroishi Site in Kuroishi, Japan (FEI 3003995201), from 2016 to 2025. The latest ended on and was classified NAI, no action indicated. FDA issued a warning letter on (CDRH).

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two of the four FOIA requests for this site have asked FDA for this inspection's 483, and one more for the records from a later inspection, ending Nov 12, 2025, that FDA's Data Dashboard doesn't list yet.

FDA’s Import Alert 89-04 Red List names this firm at this address. FDA printed no FEI on the entry. No court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.100(a) and 21 CFR 820.75(a).

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
3, 2016–2025
0 OAI · 1 VAI · 2 NAI
Inspections with 483 observations shown
2 of 3
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
4 requests, latest Sep 25, 2026
Import refusals
None under this FEI
Import alerts
89-04Red List
Matched on FDA’s firm name and street address

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record.

    2. Manufacturer Name and Address

      An individual medical device manufacturer report submitted per Form FDA 3500A format did not contain in Block D the device manufacturer's name and address.

    3. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    4. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Four requests to FDA under the Freedom of Information Act named this site, most recently on Sep 25, 2026.

  • Inspection ending Nov 12, 2025

    1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.

    1. May 1, 2026FDA-FOIA-2026-4360EIR· 1 of 8 sitesClosed
  • Inspection ending Nov 7, 2025NAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jun 18, 2026. FDAgent holds no copy that FDA posted.

    1. Sep 25, 2026FDA-FOIA-2026-9319Form 483 + EIROpen
    2. Jan 14, 2026FDA-FOIA-2026-497Form 483· 1 of 2 sitesRecords released
  • Inspection ending Jul 14, 2022VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Feb 5, 2024FDA-FOIA-2024-1255Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.

  • 21 CFR 820.75Process validation2 of 2
  • 21 CFR 820.100Corrective and preventive action2 of 2
  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
  • 21 CFR 803.52If I am a manufacturer, what information must I submit in my individual adverse event reports?1 of 2
  • 21 CFR 820.30Design controls1 of 2
  • 21 CFR 820.184Device history record1 of 2

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Import alerts

FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.

  • 89-04Red ListDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsImport alert pageAlert on fda.gov

    Date published 2026-10-19

    FDA’s entry
    AOMORI OLYMPUS CO., LTD. Kuroishi Site · 2 Chome 248-1, Okkonoki, Kuroishi, Aomori JAPAN
    FDA’s inspection record for FEI 3003995201
    AOMORI OLYMPUS CO., LTD. Kuroishi Site · 2 Chome 248-1, Okkonoki, Kuroishi, Japan

    Matched on exact firm name, street address and country. FDA printed no FEI on this entry.

    2 products

    • 77 E - - OQBronchoscope (Flexible Or Rigid)
    • 78 F - - GBUreteroscope

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003995201 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003995201. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.