Skip to content

Nipro Diabetes Systems, Inc.

3361 Enterprise Way, Miramar, Florida, United States

FEI 3003352163

FDA has one inspection on file for Nipro Diabetes Systems, Inc. in Miramar, Florida (FEI 3003352163). It ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 10 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
1, in 2009
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation - user needs and intended uses

      Design validation did not ensure that devices conform to defined user/patient needs and intended uses.

    2. Procedures

      The procedures addressing documentation of corrective and preventive action activities were not complete and implemented.

    3. Design verification - unresolved discrepancies

      Unresolved discrepancies were noted at the completion of the design verification.

    4. Design verification - documentation

      The design verification results, including methods, were not documented and filed in the design history file.

    5. Lack of procedures, or not maintained

      Document control procedures were not established, defined, and implemented.

    Show 5 more citations
    1. Quality audits - Lack of or inadequate procedures

      Procedures for conducting quality audits were not complete.

    2. Design verification acceptance criteria

      Acceptance criteria were not established, defined, documented, and complete prior to the performance of verification activities.

    3. Techniques not used

      Statistical techniques are not used for control purposes where statistical techniques are applicable.

    4. Design control - no procedures

      Procedures to control the design process of the device were not complete and implemented.

    5. Design verification - Lack of or inadequate procedures

      Design verification did not confirm that the design output meets the design input requirements.

Regulations FDA cited here

The 5 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003352163 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003352163. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.