Omron Healthcare Co., Ltd
Omuronherusukea (Kabu) Matsusakajigyosho, 1855-370; Kubo-Cho, Matsusaka, Japan
FEI 3002807674
FDA has 3 inspections on file for Omron Healthcare Co., Ltd in Matsusaka, Japan (FEI 3002807674), from 2011 to 2018. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
2 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 3, 2011–2018
- 0 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 2 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 2 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Review and evaluation for investigation
Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.
Complaints
Complaint files are not adequately maintained.
Rationale documented for no investigation
Records for complaints where no investigation was made do not include required information.
Sampling plans
Sampling plans are not based on valid statistical rationale.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 1 more citation
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.184Device history record1 of 2
- 21 CFR 820.198Complaint files1 of 2
- 21 CFR 820.250Statistical techniques1 of 2
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3002807674, one row per refused product line.
2 refusals from 2023-01-10 to 2023-01-17 · charged NOT LISTED on 2, FRNMFGREG on 2 · no FDA sample analysis
FDA firm profile- 2023-01-17
SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE74DXN
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-2364373-8/11/1
- 2023-01-10
SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE74DXN
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports799-2348825-8/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002807674 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002807674. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.