Skip to content

Northeast Laboratories, Inc.

30 Cold Spring Rd, Rocky Hill, Connecticut, United States

FEI 3001743770DUNS 086650488

FDA has 4 inspections on file for Northeast Laboratories, Inc. in Rocky Hill, Connecticut (FEI 3001743770), from 2010 to 2024. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

No warning letter, court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 211.194(a)(8).

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for analysis, through Dec 31, 2026. Its registrant is Northeast Laboratories, Inc. (D-U-N-S 086650488).

Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
4, 2010–2024
0 OAI · 2 VAI · 2 NAI
Inspections with 483 observations shown
2 of 4
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
  2. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Complete Test Data

      Laboratory records are deficient in that they do not include a complete record of all data obtained during testing.

    2. Changes to Procedures Not Reviewed, Approved

      Changes to written procedures are not drafted, reviewed and approved by the appropriate organizational unit and reviewed and approved by the quality control unit.

    3. GMP Training Frequency

      GMP training is not conducted on a continuing basis to assure that employees remain familiar with CGMP requirements applicable to them.

    4. Reference and method not stated

      Laboratory records of methods of testing used do not indicate the method and provide the reference when employing methods in recognized standard references.

    5. Identification of Person Performing Review of Lab Records

      Laboratory records are deficient in that they do not include the initials and signature of the second person reviewing the record for accuracy.

    Show 4 more citations
    1. Procedures not in writing, fully followed

      The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed.

    2. Procedure Deviations Recorded and Justified

      Deviations from written production and process control procedures are not recorded.

    3. Calibration - at intervals, written program, remedial action

      The calibration of instruments is not done at suitable intervals in accordance with an established written program.

    4. Investigations of discrepancies, failures

      There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Second person sign off

      Laboratory records do not include the initials or signature of a second person showing that the original records have been reviewed for accuracy, completeness, and compliance with established standards.

    2. Scientifically sound laboratory controls

      Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that components and drug products conform to appropriate standards of identity, strength, quality and purity.

  4. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.

FDA drug establishment registration

FDA registration search
Registered operations
Analysis
Registration runs through
Dec 31, 2026
Registrant
Northeast Laboratories, Inc.D-U-N-S 086650488

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3001743770 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3001743770. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.