FEI 3000779542

FDA has 9 inspections on file for Tissue Transplant Technology, Ltd. in San Antonio, Texas (FEI 3000779542), from 2011 to 2024. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 3 citations from this inspection, all under 21 CFR Part 1271, the rules for human cell and tissue products. They are listed below. One FOIA request has asked FDA for this inspection's 483.

FDA issued a warning letter on (CBER). No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
9, 2011–2024
0 OAI · 2 VAI · 7 NAI
Inspections with 483 observations shown
4 of 9
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest May 17, 2024
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

9 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Use of non-sterile reagents

      Reagents used in the processing and preservation of HCT/Ps were not sterile.

    2. Records maintained concurrently

      Records were not maintained concurrently with the performance of each step.

    3. Use prior to verification

      Supplies and reagents were used before they were verified to meet specifications designed to prevent the introduction, transmission, or spread of communicable disease.

  2. NAINo objectionable conditions that need follow-upBiologics · Vaccines and Allergenic Products4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigations of discrepancies, failures

      There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.

    2. Expiration date lacking

      Drug products do not bear an expiration date determined by appropriate stability data to assure they meet applicable standards of identity, strength, quality and purity at the time of use.

    3. Absence of Written Procedures

      There are no written procedures for production and process controls designed to assure that the drug products have the identity, strength, quality, and purity they purport or are represented to possess.

    4. Components withheld from use pending release

      Each lot of components is not withheld from use until the lot has been sampled, tested, examined, and released by the quality control unit.

  3. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood Products3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Deviations--evaluation, cause, corrective action

      Investigation of deviations related to core CGTP requirements did not include a review and evaluation of the deviation, efforts to determine the cause of the deviation and corrective action(s) to address the deviation and prevent recurrence.

    2. Verification/validation

      Process changes were not validated and verified to ensure the change does not cause an adverse impact elsewhere in operations.

    3. Change activities documented

      Activities related to a change to a validated process were not documented.

  4. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  5. NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
  6. NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
  7. NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
  8. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Procedures to meet core CTGP

      Procedures appropriate to meet core CGTP requirements for all steps that you perform in the manufacture of HCT/Ps were not followed.

  9. NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on May 17, 2024. It asked for a Form 483.

  • Inspection ending May 3, 2024VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. May 17, 2024FDA-FOIA-2024-4480Form 483· 1 of 2 sitesClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 9 inspections, with how many of them cited each.

Show all 10

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCBER · BiologicsFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3000779542 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3000779542. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.