Prince Medical S.A.S.
Zone Dactivite La Sente Du Moulin, 64 Rue Gutenberg, Ercuis, France
FEI 3000296634
FDA has one inspection on file for Prince Medical S.A.S. in Ercuis, France (FEI 3000296634). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH). FDA has also posted a Form 483 dated Dec 8, 2022 that FDAgent has not matched to an inspection on the Data Dashboard.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 4 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below. One FOIA request asked FDA for the 483 from a later inspection, ending Dec 8, 2022, that FDA's Data Dashboard doesn't list yet.
No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 1, in 2012
- Inspections with 483 observations shown
- 1 of 1
- 1 posted 483 not matched to an inspection
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 1 request, latest Sep 4, 2025
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Sep 4, 2025. It asked for a Form 483.
Inspection ending Dec 8, 2022
1 request. Not listed in FDA’s Data Dashboard yet. FDA has posted its Form 483.
- Sep 4, 2025FDA-FOIA-2025-10561Form 483· 1 of 2 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 4 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.75Process validation1 of 1
- 21 CFR 820.198Complaint files1 of 1
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3000296634 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3000296634. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.