Community Blood Bank of Erie County
2646 Peach St, Erie, Pennsylvania, United States
FEI 2572105
FDA has 6 inspections on file for Community Blood Bank of Erie County in Erie, Pennsylvania (FEI 2572105), from 2009 to 2024. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
No warning letter, court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 6, 2009–2024
- 0 OAI · 3 VAI · 3 NAI
- Inspections with 483 observations shown
- 4 of 6
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Concurrent documentation
Records are not concurrently maintained with the performance of each significant step in the storage of each unit of blood and blood components so that all steps can be clearly traced.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the processing of blood and blood components for allogeneic transfusion were not always.
Thorough investigations
Failure to perform a thorough investigation of an unexplained discrepancy.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Written calibration / inspection records not kept
Records of the inspections of automatic, mechanical or electronic equipment, including computers or related systems are not maintained.
Thorough investigations
Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.
Equipment observed, standardized, calibrated
Equipment used in the collection of blood and blood components is not calibrated on a regularly scheduled basis as prescribed in the SOP Manual.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Donor suitability by means of medical history
Failure to always determine donor suitability on the day of collection by means of test for hemoglobin level.
Testing - inspection
Failure to visually inspect blood during storage and immediately prior to issue for abnormal color, physical appearance, or indication or suspicion of microbial contamination.
Concurrent documentation
Records are not concurrently maintained with the performance of each significant step in the processing and storage of each unit of blood and blood components so that all steps can be clearly traced.
Changes to Procedures Not Reviewed, Approved
Changes to written procedures are not drafted, reviewed and approved by the appropriate organizational unit.
Equipment observed, standardized, calibrated
Equipment used in the collection and processing of blood and blood components is not observed and calibrated on a regularly scheduled basis as prescribed in the SOP Manual.
Show 2 more citations
Thorough investigations
Failure to perform a thorough investigation of an unexplained discrepancy.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection and processing of blood and blood components for homologous transfusion are not always maintained and followed.
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood Products6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Adverse reaction - Maintenance of Reports
Failure to maintain reports of complaints of adverse reactions regarding each unit of blood or blood product arising as a result of blood collection.
Concurrent documentation
Records are not concurrently maintained with the performance of each significant step in the processing, compatibility testing, and storage of each unit of blood and blood components so that all steps can be clearly traced.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed.
Authority lacking to review records, investigate errors
The quality control unit lacks authority to review production records to assure that no errors have occurred and fully investigate errors that have occurred.
Qualifications of responsible personnel
The personnel responsible for the collection, processing, compatibility testing, storage, and distribution of blood or blood components are not adequate in training and experience, including professional training as necessary to assure competent performance of their assigned functions, and to ensure that the final product has the safety, purity, potency, identity and effectiveness it purports or is represented to possess.
Show 1 more citation
Equipment calibration frequency
Equipment used in the processing of blood and blood components is not observed, standardized and calibrated with at least the frequency required.
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 606.60Equipment3 of 4
- 21 CFR 606.100Standard operating procedures3 of 4
- 21 CFR 606.160Records3 of 4
- 21 CFR 211.22Responsibilities of quality control unit1 of 4
- 21 CFR 211.68Automatic, mechanical, and electronic equipment1 of 4
- 21 CFR 211.100Written procedures; deviations1 of 4
- 21 CFR 606.20Personnel1 of 4
- 21 CFR 606.170Adverse reaction file1 of 4
Show all 10
- 21 CFR 640.31 of 4
- 21 CFR 640.5Testing the blood1 of 4
Other FEIs recorded as Community Blood Bank of Erie County
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Community Blood Bank of Erie CountyFEI 3002965653Bradford, Pennsylvania, United States
Last inspected 2011-10-15NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2572105 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2572105. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.