Skip to content

Vynacron Dental Resins, Inc.

335 Bismark Rd Unit A2, Jackson, New Jersey, United States

FEI 2242866

FDA has 6 inspections on file for Vynacron Dental Resins, Inc. in Jackson, New Jersey (FEI 2242866), from 2011 to 2026. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 8 citations from this inspection. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.

The most recent OAI inspection ended on Jan 21, 2025. No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
6, 2011–2026
1 OAI · 0 VAI · 5 NAI
Inspections with 483 observations shown
3 of 6
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Aug 13, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 8.5.3

      Preventive action procedures not documented

      Preventive action procedures describing the requirements for determining potential nonconformities and their causes, evaluating the need for action to prevent occurrence of potential nonconformities, planning and documenting action needed and implementing such action, verifying that the action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of the preventive action taken have not been adequately documented.

    2. ISO 13485:2016 Clause 4.1.2

      Risk-based approach not applied to QMS processes

      A risk-based approach to the control of the appropriate processes needed for the quality management system has not been applied.

    3. ISO 13485:2016 Clause 7.3.3

      Design and development inputs records not maintained

      Records for design and development inputs related to product requirements were not maintained.

    4. ISO 13485:2016 Clause 8.5.2

      Corrective action procedures not documented

      Corrective action procedures defining the requirements for reviewing nonconformities, verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of corrective action taken have not been adequately documented.

    5. ISO 13485:2016 Clause 7.6(a)

      Measuring equipment not adequately calibrated or verified

      Measuring equipment was not calibrated and verified at specified intervals or prior to use against measurements traceable to international or national measurement standards.

    Show 3 more citations
    1. ISO 13485:2016 Clause 7.3.4

      Records of design and development outputs not maintained

      Records of design and development outputs were not maintained.

    2. ISO 13485:2016 Clause 5.6.1

      Management review procedures not documented

      Procedures for management review have not been documented.

    3. ISO 13485:2016 Clause 8.2.2

      Complaint handling procedures lack required elements

      Procedures for timely complaint handling do not adequately include requirements and responsibilities for determining the need to report the information to FDA, handling of complaint-related product and determining the need to initiate corrections or corrective actions.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Lack of Written MDR Procedures

      Written MDR procedures have not been developed, maintained and implemented.

    4. Procedures not adequately established or maintained

      Document control procedures have not been adequately established and maintained.

  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  5. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  6. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Aug 13, 2026. Each one asked for a Form 483.

  • Inspection ending Jun 25, 2026NAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 31, 2026. FDAgent holds no copy that FDA posted.

    1. Aug 13, 2026FDA-FOIA-2026-7847Form 483· 1 of 5 sitesRecords released
  • Inspection ending Jan 21, 2025OAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Mar 18, 2025FDA-FOIA-2025-2494Form 483· 1 of 2 sitesClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 5 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 6 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2242866 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2242866. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.