Skip to content

Scantibodies Laboratory, Inc.

9336 Abraham Way, Santee, California, United States

FEI 2020808

FDA has 5 inspections on file for Scantibodies Laboratory, Inc. in Santee, California (FEI 2020808), from 2011 to 2019. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

One FOIA request asked FDA for the 483 from a later inspection, ending Jul 2, 2026, that FDA's Data Dashboard doesn't list yet.

The most recent OAI inspection ended on May 24, 2013. No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
5, 2011–2019
1 OAI · 1 VAI · 3 NAI
Inspections with 483 observations shown
2 of 5
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Sep 9, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  2. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the collection, processing, storage, and distribution of blood and blood components for further manufacturing purposes are not always followed.

    2. Written consent

      Failure to obtain written consent of prospective Source Plasma donors.

    3. Labeling changes

      Failure to receive FDA approval before distributing the product with labeling changes.

    4. Record review prior to release

      All records pertinent to a lot or unit were not reviewed before the release or distribution of a lot or unit of final product.

    5. Physician review of four month sample

      Failure of the physician or physician substitute to review and sign the plasma or serum protein electrophoresis data and collection records within 21 days after the sample was drawn.

    Show 3 more citations
    1. Donor reaction

      Failure of the donor record to contain a full explanation of a donor reaction, while on the plasmapheresis premises or reported to the center after the donor has left the premises, including the measures taken to assist the donor and the outcome of the incident.

    2. Qualifications of responsible personnel

      The personnel responsible for the collection, processing, storage, and distribution of blood or blood components are not adequate in training and experience, including professional training as necessary to assure competent performance of their assigned functions, and to ensure that the final product has the safety, purity, potency, identity and effectiveness it purports or is represented to possess.

    3. Donor identification system

      Failure to establish a donor identification system that positively identifies each donor and relates such donor directly to his blood and its components as well as to his accumulated records and laboratory data.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionBiologics · Blood and Blood ProductsCitations not published
  4. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  5. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Not approved or obsolete document retrieval

      Documents that were not approved were observed at a location where they are being used.

    2. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been adequately established.

    3. Documentation

      Acceptance activities were not adequately maintained as part of the device history record.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Sep 9, 2026. It asked for a Form 483.

  • Inspection ending Jul 2, 2026

    1 request. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on Sep 28, 2026. FDAgent holds no copy that FDA posted.

    1. Sep 9, 2026FDA-FOIA-2026-8735Form 483· 1 of 4 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 5 inspections, with how many of them cited each.

Show all 10

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2020808 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2020808. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.