Koven Technology, Inc.
625 Maryville Centre Dr Ste 300, Saint Louis, Missouri, United States
FEI 1937397
FDA has 3 inspections on file for Koven Technology, Inc. in Saint Louis, Missouri (FEI 1937397), from 2017 to 2026. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA posted the Form 483 from that inspection. FDA published 4 citations from this inspection. They are listed below. Four FOIA requests have asked FDA for this inspection's 483.
No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Form 483 PDF ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 3, 2017–2026
- 1 OAI · 0 VAI · 2 NAI
- Inspections with 483 observations shown
- 2 of 3
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 4 requests, latest Sep 14, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesPosted 483 · 4 citations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.4.1
Criteria for the evaluation and selection of suppliers not established
Criteria for the evaluation and selection of suppliers were not adequately established.
- ISO 13485:2016 Clause 7.3.9
Significance of design and development changes not determined
The significance of the design and development change to safety and regulatory requirements for the medical device and its intended use was not determined.
- ISO 13485:2016 Clause 7.1
Required elements not determined in planning product realization
In product realization planning your firm did not adequately determine, as appropriate, quality objectives and requirements for the product.
Records for complaints involving failure to meet specifications not maintained
Records of the review, evaluation, and investigation for any complaints involving the possible failure of a device, labeling, or packaging to meet any of its specifications were not adequately maintained.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Sep 14, 2026. Each one asked for a Form 483.
Inspection ending Feb 6, 2026OAI
4 requests. In the inspection history above. FDA has posted its Form 483.
- Sep 14, 2026FDA-FOIA-2026-8874Form 483 + EIROpen
- Aug 13, 2026FDA-FOIA-2026-7836Form 483Open
- May 19, 2026FDA-FOIA-2026-5147Form 483· 1 of 4 sitesRecords released
- Apr 27, 2026FDA-FOIA-2026-4185Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 2 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.35Control of records1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1937397 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1937397. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.