Koven Technology, Inc. Form 483, February 2026
FDA issued this Form 483 to Koven Technology, Inc. in Saint Louis, Missouri (FEI 1937397) at the end of a device inspection that ended February 6, 2026. FDA classified the inspection Official Action Indicated (OAI) and issued the facility a warning letter on July 21, 2026.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Nov 6, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Feb 6, 2026
- Devices · Compliance: Devices
- FDA's outcome
- OAI
- Official Action Indicated
- Posted by FDA
- Nov 6, 2026
- 273 days after the inspection
FDA's 4 citations from this inspection
FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.
- ISO 13485:2016 Clause 7.4.1
Criteria for the evaluation and selection of suppliers not established
Criteria for the evaluation and selection of suppliers were not adequately established.
- ISO 13485:2016 Clause 7.3.9
Significance of design and development changes not determined
The significance of the design and development change to safety and regulatory requirements for the medical device and its intended use was not determined.
- ISO 13485:2016 Clause 7.1
Required elements not determined in planning product realization
In product realization planning your firm did not adequately determine, as appropriate, quality objectives and requirements for the product.
Records for complaints involving failure to meet specifications not maintained
Records of the review, evaluation, and investigation for any complaints involving the possible failure of a device, labeling, or packaging to meet any of its specifications were not adequately maintained.
After this inspection
FDA's compliance actions for this FEI dated after this inspection and before its next one. FDAgent links a letter when FDA's case number ties it to this facility.
- Warning LetterJul 21, 2026Koven Technologies, Inc. warning letter, July 21, 2026CDRH · 165 days after this inspection ended
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.