Montrose Memorial Hospital Blood Bank
800 S 3rd St, Montrose, Colorado, United States
FEI 1774501
FDA has 4 inspections on file for Montrose Memorial Hospital Blood Bank in Montrose, Colorado (FEI 1774501), from 2010 to 2017. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Sep 17, 2010. No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 606.100(b) and 21 CFR 640.3(b)(6).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 4, 2010–2017
- 1 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 2 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Donor suitability procedures not followed
Failure to follow standard procedures and methods for determining the suitability of a donor as a source of blood.
Second person sign off
Laboratory records do not include the initials or signature of a second person showing that the original records have been reviewed for accuracy, completeness, and compliance with established standards.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection, processing, storage, and distribution of blood and blood components for homologous transfusion are not always maintained and followed.
Criteria for suitability of reissue of returned blood
The standard operating procedure fails to include a written description of the criteria for determining whether returned blood is suitable for reissue.
Concurrent documentation
Records are not concurrently maintained with the performance of each significant step in the collection of each unit of blood and blood components so that all steps can be clearly traced.
Show 3 more citations
Donor qualifying tests & measurements
The standard operating procedure fails to include written descriptions of methods for performing donor qualifying tests and measurements, including minimum and maximum values for a test or procedure when a factor in determining acceptability.
Qualifications of donor - disease transmissible by blood
Failure to determine whether each donor is free from any disease transmissible by blood as determined by history and examinations.
Notification w/in 8 weeks
Failure to make reasonable attempts to notify the donor within 8 weeks after determining that the donor is deferred or determined not to be suitable for donation.
- OAISignificant objectionable conditions; FDA recommended regulatory actionBiologics · Blood and Blood Products8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Labeling of blood products unsuitable for transfusion
Failure to prominently label blood and blood components (except for recovered plasma) determined to be unsuitable for transfusion with "NOT FOR TRANSFUSION" and the reason the unit is considered unsuitable.
Maintenance - tracing
Records are not maintained in a manner which allows steps in the manufacture and distribution of product to be traced.
Following manufacturer's instructions
Failure to use supplies and reagents in a manner consistent with instructions provided by the manufacturer.
Adverse reaction - fatality
A confirmed, fatal complication of transfusion was not reported as soon as possible and submitted in writing within 7 days after the fatality to the Director, Office of Compliance, Center for Biologics Evaluation and Research by the collecting facility in the event of a donor reaction and facility that performed the compatibility tests in the event of a transfusion reaction.
Provide space for quarantine of unsuitable for use items
Failure to provide adequate space for the quarantine and storage of products and not suitable for use.
Show 3 more citations
Donor suitability by means of medical history
Failure to always determine donor suitability on the day of collection by means of medical history and physical examination.
Qualifications of donor - disease transmissible by blood
Failure to determine whether each donor is free from any disease transmissible by blood as determined by history and examinations.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection, processing, and distribution of blood and blood components for homologous transfusion and autologous transfusion are not always maintained and followed.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 606.100Standard operating procedures2 of 2
- 21 CFR 640.32 of 2
- 21 CFR 211.194Laboratory records1 of 2
- 21 CFR 600.12Records1 of 2
- 21 CFR 606.40Facilities1 of 2
- 21 CFR 606.65Supplies and reagents1 of 2
- 21 CFR 606.121Container label1 of 2
- 21 CFR 606.160Records1 of 2
Show all 10
- 21 CFR 606.170Adverse reaction file1 of 2
- 21 CFR 630.40Requirements for notifying deferred donors1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1774501 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1774501. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.