Central Shared Services, LLC.
5405 S 500 E Ste 105, Ogden, Utah, United States
FEI 1771298
FDA has 3 inspections on file for Central Shared Services, LLC. in Ogden, Utah (FEI 1771298), from 2009 to 2014. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 606.100(b), 21 CFR 606.100(c) and 21 CFR 606.171.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2009–2014
- 0 OAI · 3 VAI · 0 NAI
- Inspections with 483 observations shown
- 3 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Blood & Blood Components
Failure to maintain temperatures during shipment of Red Blood Cells between 1 and 10 ?C and Whole Blood as required.
Criteria for plasmapheresis donor eligibility
Failure of a plasmapheresis donor to be presented the informed consent.
Thorough investigations
Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection, processing, storage, and distribution of blood and blood components for homologous transfusion are not always maintained and followed.
Controlling storage temperatures
The standard operating procedure fails to include a written description of the storage temperatures and methods of controlling storage temperatures for all blood products and reagents
Show 1 more citation
Biological product deviation report
Failure to submit a biological product deviation report within 45 days from the date you acquired information suggesting that a reportable event occurred.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection, processing, and storage of blood and blood components for homologous transfusion and further manufacturing purposes are not always maintained and followed.
Required records
Failure to maintain storage and distribution records.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate sampling plans designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity.
Following manufacturer's instructions
Failure to use supplies and reagents in a manner consistent with instructions provided by the manufacturer.
Biological product deviation report
Failure to submit a biological product deviation report within 45 days from the date you acquired information suggesting that a reportable event occurred.
Show 1 more citation
Equipment observed, standardized, calibrated
Equipment used in the of blood and blood components is not observed and calibrated on a regularly scheduled basis as prescribed in the SOP Manual.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Thorough investigations
Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 606.100Standard operating procedures3 of 3
- 21 CFR 606.171Reporting of product deviations by licensed manufacturers, unlicensed registered blood establishments, and transfusion services2 of 3
- 21 CFR 211.160General requirements1 of 3
- 21 CFR 600.15Temperatures during shipment1 of 3
- 21 CFR 606.60Equipment1 of 3
- 21 CFR 606.65Supplies and reagents1 of 3
- 21 CFR 606.160Records1 of 3
- 21 CFR 640.31Eligibility of donors1 of 3
Other FEIs recorded as Central Shared Services, LLC.
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Central Shared Services, LLC.FEI 3003832552Provo, Utah, United States
Last inspected 2014-11-10NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1771298 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1771298. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.