Inova Health Care Services

45745 Nokes Blvd Ste 160, Sterling, Virginia, United States

FEI 1177788

FDA has 8 inspections on file for Inova Health Care Services in Sterling, Virginia (FEI 1177788), from 2009 to 2024. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 8 citations from this inspection. They are listed below. One of the three FOIA requests for this site has asked FDA for this inspection's 483.

The most recent OAI inspection ended on Feb 13, 2012. FDA issued a warning letter on (CBER). No court action or import refusal is filed under this FEI.

Six regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
8, 2009–2024
1 OAI · 6 VAI · 1 NAI
Inspections with 483 observations shown
6 of 8
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
3 requests, latest Jan 22, 2025
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Ineligible Donor

      Blood or blood components were collected from a donor found to be deferred under the regulations prior to collection.

    2. Concurrent documentation

      Records are not concurrently maintained with the performance of each significant step in the processing of each unit of blood and blood components so that all steps can be clearly traced.

    3. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the processing and distribution of blood and blood components for allogeneic transfusion, autologous transfusion and further manufacturing purposes were not always maintained.

    4. Who must report - manufacturer

      Failure to submit a biological deviation reports.

    5. Notify Consignees [Collecting Establishment]

      Failure to notify consignees of the results of further testing for HIV within 45 calendar days after the donor tests reactive for evidence of HIV infection under the regulations.

    Show 3 more citations
    1. Procedures not in writing, fully followed

      The responsibilities and procedures applicable to the quality control unit are not fully followed.

    2. Notify Consignees [Collecting Establishment]

      Failure to notify consignees to quarantine all previously collected in-date blood and blood components identified under the regulations (blood and blood components collected twelve months and less before the donor's most recent nonreactive screening tests or twelve months and less before the donor's reactive direct viral detection test) if intended for use for further manufacture into injectable products.

    3. Notify Consignees [Collecting Establishment]

      Failure to notify consignees of the results of further testing for HCV and the results of the reactive screening test within 45 calendar days after the donor tests reactive for evidence of HCV infection under the regulations.

  2. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Proof Of Identity

      Failure to obtain from the donor on the day of donation a postal address where the donor may be contacted for 8 weeks after donation.

    2. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the collection, processing and distribution of blood and blood components for allogeneic transfusion and further manufacturing purposes were not always maintained and followed.

    3. Form and manner of educational material

      Failure to present educational material to the donor in an appropriate form.

  3. NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene Therapies2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Thorough investigations

      Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.

    2. Record review prior to release

      All records pertinent to a lot or unit were not reviewed before the release or distribution of a lot or unit of final product.

  4. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsCitations not published
  5. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsCitations not published
  6. OAISignificant objectionable conditions; FDA recommended regulatory actionBiologics · Blood and Blood Products10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Who must report - manufacturer

      Failure to submit biological deviation reports.

    2. Qualifications of donor - blood pressure

      Each donor was not in good health as indicated in part by systolic and diastolic blood pressures which were not within normal limits and the examining physician has not determined that this individual with blood pressures outside these limits is otherwise qualified as a donor.

    3. Written methods for investigating adverse reactions

      The standard operating procedure fails to include a written description of the procedures for investigating adverse donor and recipient reactions.

    4. Person performing, test results, interpretation

      Records fail to be as detailed as necessary so as to provide a complete history of the work performed.

    5. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the collection and processing of blood and blood components for homologous transfusion, autologous transfusion, and further manufacturing purposes are not always maintained and followed.

    Show 5 more citations
    1. Determination of lot numbers and supplies

      Appropriate records are not available to determine the lot numbers of supplies used for specific units of the final product.

    2. Thorough investigations

      Failure to make a record of the conclusions and follow-up of an unexplained discrepancy.

    3. Cleaning/maintenance records not kept

      Records are not kept for the maintenance of equipment.

    4. Equipment observed, standardized, calibrated

      Equipment used in the processing of blood and blood components is not observed, standardized, and calibrated on a regularly scheduled basis as prescribed in the SOP Manual.

    5. Identity of major equipment and lines used

      Batch production and control records do not include the identity of individual major equipment used for each batch of drug product produced.

  7. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Adverse Reaction- Reports of Investigations

      Written reports of investigations of adverse reactions, including conclusions and follow up, are not prepared and maintained.

    2. Maintain and clean equipment

      Failure to maintain equipment used in the storage of blood and blood products in a clean and orderly manner.

    3. Person performing, test results, interpretation

      Records fail to be as detailed as necessary so as to provide a complete history of the work performed.

    4. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the collection, processing, and storage of blood and blood components for homologous transfusion and further manufacturing purposes are not always maintained and followed.

    5. Procedures not in writing, fully followed

      The responsibilities and procedures applicable to the quality control unit are not fully followed.

    Show 5 more citations
    1. Equipment calibration frequency

      Equipment used in the processing and storage and distribution of blood and blood components is not observed, standardized and calibrated with at least the frequency required.

    2. Concurrent documentation

      Records are not concurrently maintained with the performance of each significant step in the processing of each unit of blood and blood components so that all steps can be clearly traced.

    3. Adverse reaction - Maintenance of Reports

      Failure to maintain reports of complaints of adverse reactions regarding each unit of blood or blood product arising as a result of blood collection.

    4. Required records

      Failure to maintain storage and distribution records.

    5. Provide space for examination

      Failure to provide adequate space for private examinations of individuals to determine their suitability as blood donors.

  8. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Arm preparation

      The phlebotomy site is not prepared by a method that gives maximum assurance of a sterile container of Whole Blood.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on Jan 22, 2025. Each one asked for a Form 483.

  • Inspection ending Sep 4, 2024VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Nov 1, 2024FDA-FOIA-2024-9657Form 483Closed
  • Inspection ending Aug 7, 2019VAI

    2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Jan 22, 2025FDA-FOIA-2025-633Form 483Closed
    2. Nov 27, 2023FDA-FOIA-2023-10576Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 14 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 8 inspections, with how many of them cited each.

Show all 14

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCBER · BiologicsFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1177788 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1177788. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.