Gyrus ACMI, Inc.
2925 Appling Rd, Bartlett, Tennessee, United States
FEI 1037007
FDA has 4 inspections on file for Gyrus ACMI, Inc. in Bartlett, Tennessee (FEI 1037007), from 2010 to 2022. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 4, 2010–2022
- 0 OAI · 1 VAI · 3 NAI
- Inspections with 483 observations shown
- 1 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Design validation - documentation
The results of design validation, including identification of the design, method(s), the date, and the individual(s) performing validation, were not documented in the design history file.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Regulations FDA cited here
The 3 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 4 inspections.
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.90Nonconforming product1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
Other FEIs recorded as Gyrus ACMI, Inc.
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Gyrus ACMI, Inc.FEI 3011050570Brooklyn Park, Minnesota, United States
Last inspected 2025-11-12OAI
Gyrus ACMI, Inc.FEI 3003696010Littleton, Massachusetts, United States
Last inspected 2023-11-02VAI
Gyrus ACMI, Inc.FEI 3004450998Redmond, Washington, United States
Last inspected 2018-04-17VAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1037007 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1037007. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.