Eurofins Donor & Product Testing LLC
6933 S Revere Pkwy, Centennial, Colorado, United States
FEI 1000477683
FDA has 10 inspections on file for Eurofins Donor & Product Testing LLC in Centennial, Colorado (FEI 1000477683), from 2010 to 2024. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.
No warning letter, court action or import refusal is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 10, 2010–2024
- 0 OAI · 6 VAI · 4 NAI
- Inspections with 483 observations shown
- 5 of 10
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
10 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene TherapiesCitations not published
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood Products5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures for cleaning, sanitation
Procedures for facility cleaning and sanitation were not followed.
State of repair
Facilities were not maintained in a good state of repair.
Calibration procedures and schedules (general)
The procedures for the calibration of equipment used for measuring were not established and defined.
Procedures to meet core CTGP
Procedures appropriate to meet core CGTP requirements for all steps that you perform in the manufacture of HCT/Ps were not followed.
Records incomplete
Records did not identify the person performing the work and were not detailed as necessary to provide a complete history of work performed.
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Records incomplete
Records did not identify the person performing the work , did not show the dates of entries and were not detailed as necessary to provide a complete history of work performed .
Corrective actions
Appropriate corrective actions were not taken related to the inspections of environmental control systems.
Manufacturer instructions not followed
Testing for communicable disease agents was not performed in accordance with the manufacturer's instructions.
Procedures to meet core CTGP
Procedures appropriate to meet core CGTP requirements for all steps that you perform in the manufacture of HCT/Ps were not established , defined , implemented and followed .
Verification/validation
Process changes were not validated and verified to ensure the change does not cause an adverse impact elsewhere in operations.
Show 5 more citations
Aseptic processing---cleaning, disinfecting
Environmental controls do not provide for adequate cleaning of rooms and equipment to ensure aseptic processing.
Size, construction, location--prevent mix-ups etc.
Facilities were not of suitable to prevent contamination of HCT/Ps with communicable disease agents .
Validation & approval--established procedures
Processes with results which could not be fully verified by inspection and tests, were not validated and approved according to established procedures.
Calibration procedures and schedules (general)
The procedures for the calibration of equipment used for measuring were not established and defined .
Procedures inadequate
Procedures for the cleaning , sanitizing and maintenance of equipment were not maintained , defined , documented and followed .
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene TherapiesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Manufacturer instructions not followed
Testing for communicable disease agents was not performed in accordance with the manufacturer's instructions.
Procedures to meet core CTGP
Procedures appropriate to meet core CGTP requirements for all steps that you perform in the manufacture of HCT/Ps were not followed.
Actions verified, short and long term solutions
Corrective actions relating to core CGTP requirements have not been verified to ensure effectiveness and compliance with CGTP and did not include both short term corrective actions to address the immediate deficiency and long term corrective actions to prevent recurrence.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Trained or re-trained as necessary
Personnel have not been trained to adequately perform their assigned responsibilities.
Procedures to meet core CTGP
Procedures appropriate to meet core CGTP requirements for all steps that you perform in the manufacture of HCT/Ps were not followed.
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Thorough investigations
Failure to make a record of the conclusions and follow-up of a failure of a lot or unit to meet any of its specifications.
Program is appropriate for the HCT/Ps
A quality program appropriate for the HCT/Ps manufactured and manufacturing steps performed has not been maintained, defined, documented, and followed.
Person performing, test results, interpretation
Records fail to include dates of the various entries and be as detailed as necessary so as to provide a complete history of the work performed.
Following manufacturer's instructions
Failure to use supplies and reagents in a manner consistent with instructions provided by the manufacturer.
QC procedures for supplies and reagents
The standard operating procedure fails to include a written description of the quality control procedures for supplies and reagents employed in processing.
Show 2 more citations
SOPs not followed / documented
Written production and process control procedures are not followed in the execution of production and process control functions.
Determination of lot numbers and supplies
Appropriate records are not available to determine the lot numbers of supplies and reagents used for specific lots of the final product.
Regulations FDA cited here
The 14 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 10 inspections, with how many of them cited each.
- 21 CFR 1271.180Procedures4 of 5
- 21 CFR 1271.80What are the general requirements for donor testing?2 of 5
- 21 CFR 1271.160Establishment and maintenance of a quality program2 of 5
- 21 CFR 1271.190Facilities2 of 5
- 21 CFR 1271.200Equipment2 of 5
- 21 CFR 1271.270Records2 of 5
- 21 CFR 211.100Written procedures; deviations1 of 5
- 21 CFR 606.65Supplies and reagents1 of 5
Show all 14
- 21 CFR 606.100Standard operating procedures1 of 5
- 21 CFR 606.160Records1 of 5
- 21 CFR 1271.170Personnel1 of 5
- 21 CFR 1271.195Environmental control and monitoring1 of 5
- 21 CFR 1271.225Process changes1 of 5
- 21 CFR 1271.230Process validation1 of 5
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000477683 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000477683. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.