PC Marketing, Inc.
1040 Wilt Ave, Ridgefield, New Jersey, United States
FEI 1000122259
FDA has 2 inspections on file for PC Marketing, Inc. in Ridgefield, New Jersey (FEI 1000122259), from 2017 to 2019. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 4 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
The most recent OAI inspection ended on Jan 19, 2017. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 803.40(b) and 21 CFR 820.198(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2017–2019
- 1 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Aug 3, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Complaints
Complaint files are not adequately maintained.
Not approved or obsolete document retrieval
Documents that were not approved were observed at a location where they could be used.
Report of malfunction likely to cause death or injury
The importer failed to submit a report to the manufacturer on FDA Form 3500A within 30 days concerning information that one of the devices marketed by the importer has malfunctioned and that such device or a similar device marketed by the importer would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Complaints
Complaint files are not adequately maintained.
Reproduction of labels
The users' instructions for the sunlamp product did not contain a reproduction of the required label(s) prominently displayed at the beginning of the instructions.
Quality System Procedures
Quality system procedures and instructions have not been established.
Lack of or inadequate organizational structure
The organizational structure has not been established and maintained to ensure that devices are in accordance with 21 CFR 820.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been established.
Show 3 more citations
Foreign complaint files
The manufacturer's formally designated complaint unit is located outside of the United States and complaint records are not reasonably accessible in the United States.
Report of malfunction likely to cause death or injury
The importer failed to submit a report to the manufacturer on FDA Form 3500A within 30 days concerning information that one of the devices marketed by the importer has malfunctioned and that such device or a similar device marketed by the importer would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Lack of Written MDR Procedures
Written MDR procedures have not been developed and implemented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Aug 3, 2026. It asked for a Form 483.
Inspection ending Jan 19, 2017OAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 20, 2026. FDAgent holds no copy that FDA posted.
- Aug 3, 2026FDA-FOIA-2026-7444Form 483Records released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 803.40If I am an importer, what reporting requirements apply to me?2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 2
- 21 CFR 1040.20Sunlamp products and ultraviolet lamps intended for use in sunlamp products1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000122259 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000122259. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.