21 CFR 606.145: Control of bacterial contamination of platelets
FDA's inspection data lists 4 Form 483 citations of 21 CFR 606.145 at 3 facilities, from April 2018 to June 15, 2023.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 4
- At 3 inspections
- Facilities cited
- 3
- Distinct FEIs
- Most recent citation
- Jun 15, 2023
- Clinical Laboratories of Hawaii, LLP
- Warning letters
- 0
- State a violation of it
The regulation
The current text of 21 CFR 606.145 from the Electronic Code of Federal Regulations.
§ 606.145 Control of bacterial contamination of platelets.
(a) Blood collection establishments and transfusion services must assure that the risk of bacterial contamination of platelets is adequately controlled using FDA approved or cleared devices or other adequate and appropriate methods found acceptable for this purpose by FDA.
(b) In the event that a blood collection establishment identifies platelets as bacterially contaminated, that establishment must not release for transfusion the product or any other component prepared from the same collection, and must take appropriate steps to identify the organism.
(c) In the event that a transfusion service identifies platelets as bacterially contaminated, the transfusion service must not release the product and must notify the blood collection establishment that provided the platelets. The transfusion service must take appropriate steps to identify the organism; these steps may include contracting with the collection establishment or a laboratory to identify the organism. The transfusion service must further notify the blood collection establishment either by providing information about the species of the contaminating organism when the transfusion service has been able to identify it, or by advising the blood collection establishment when the transfusion service has determined that the species cannot be identified.
(d) In the event that a contaminating organism is identified under paragraph (b) or (c) of this section, the collection establishment's responsible physician, as defined in § 630.3(i) of this chapter, must determine whether the contaminating organism is likely to be associated with a bacterial infection that is endogenous to the bloodstream of the donor, in accordance with a standard operating procedure developed under § 606.100(b)(22). This determination may not be further delegated.
21 CFR Part 606: Current Good Manufacturing Practice for Blood and Blood Components, Subpart HRead on eCFR
Citations by year
Form 483 citations of 21 CFR 606.145 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 2 in 2019
| Year | Form 483 citations |
|---|---|
| 2018 | 1 |
| 2019 | 2 |
| 2023 | 1 |
How FDA words this citation
FDA's short descriptions and citation text on these 4 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Control risk of contamination1 citations
- Infection in bloodstream1 citations
- Release of contaminated product1 citations
- Tx service release of product, notification, ID1 citations
Citation text
- “Failure of the responsible physician to determine whether an organism identified in a bacterially contaminated platelet is likely to be associated with a bacterial infection that is endogenous to the bloodstream of the donor.”1 citations
- “Failure of the transfusion service that identified platelets as bacterially contaminated to notify the blood collection establishment that provided the platelets and take appropriate steps to identify the organism.”1 citations
- “Failure to assure that the risk of bacterial contamination of platelets was adequately controlled by using.”1 citations
- “Failure to prevent the release for transfusion of any other component prepared from the same collection as a platelet product identified as bacterially contaminated.”1 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 606.145, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- Clinical Laboratories of Hawaii, LLP2023-06-15 · FEI 2972911Tx service release of product, notification, ID
Often cited at the same inspection
Of the 3 inspections where FDA cited 21 CFR 606.145, the share where it also cited each of these sections.
- 21 CFR 606.100Standard operating procedures3 (100%)
- 21 CFR 606.160Records2 (67%)
- 21 CFR 211.22Responsibilities of quality control unit1 (33%)
- 21 CFR 606.60Equipment1 (33%)
- 21 CFR 606.65Supplies and reagents1 (33%)
- 21 CFR 610.40Test requirements1 (33%)
Ask FDAgent about 21 CFR 606.145
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 606.145; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.