Frontier Biologics, LLC warning letter, November 1, 2024
FDA's Center for Biologics Evaluation and Research issued this warning letter to Frontier Biologics, LLC of Waco, TX on November 1, 2024 and posted it on November 19, 2024. It states 6 violations, citing 21 CFR 211.113(b), 211.100(a), 211.42(c)(10)(iv), 211.42(c)(10)(v), 211.192 and 211.166(a).
Source: FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Read the full letter on fda.gov. FDAgent is independent of FDA.
- Issued
- Nov 1, 2024
- By CBER
- Posted by FDA
- Nov 19, 2024
- 18 days after issue
- Subject
- Unapproved New Drug/Unlicensed Biological Product Violations
- As FDA lists it
- FDA reference
- MARCS-CMS 686059
- FDA's case number
Violations FDA states in this letter
Each numbered violation sentence that cites 21 CFR, quoted exactly from the letter. The full letter below explains each one. A check means FDA's citation data also lists that section for the facility's March 8, 2024 inspection.
- 1
“Failure to establish and follow appropriate written procedures designed to prevent microbiological contamination of drug products purporting to be sterile, including procedures for validation of all aseptic and sterilization processes, as required by 21 CFR 211.113(b).”
21 CFR 211.113Control of microbiological contamination1,405 Form 483 citations · 90 warning letters✓ Cited at the March 8, 2024 inspection - 2
“Failure to establish written procedures for production and process controls designed to assure that the drug products have the identity, strength, quality, and purity they purport or are represented to possess, as required by 21 CFR 211.100(a).”
21 CFR 211.100Written procedures; deviations2,673 Form 483 citations · 175 warning letters✓ Cited at the March 8, 2024 inspection - 3
“Aseptic processing areas are deficient regarding the system for environmental monitoring to prevent contamination, as required by 21 CFR 211.42(c)(10)(iv).”
21 CFR 211.42Design and construction features1,661 Form 483 citations · 81 warning letters✓ Cited at the March 8, 2024 inspection - 4
“Aseptic processing areas are deficient regarding the system for cleaning and disinfecting the room and equipment to produce aseptic conditions, as required by 21 CFR 211.42(c)(10)(v).”
21 CFR 211.42Design and construction features1,661 Form 483 citations · 81 warning letters✓ Cited at the March 8, 2024 inspection - 5
“Written records are not always made of investigations into unexplained discrepancies or the failure of a batch or any of its components to meet specifications whether or not the batch has already been distributed , as required by 21 CFR 211.192.”
21 CFR 211.192Production record review2,812 Form 483 citations · 147 warning letters✓ Cited at the March 8, 2024 inspection - 6
“Failure to establish and follow a written testing program designed to assess the stability characteristics of drug products and to use results of such stability testing to determine appropriate storage conditions and expiration dates, as required by 21 CFR 211.166(a).”
21 CFR 211.166Stability testing1,733 Form 483 citations · 94 warning letters✓ Cited at the March 8, 2024 inspection
The facility
FDA's compliance action record for this letter (case 686059) names one facility. Each facility's FDA record on FDAgent has its full inspection history.
Inspections
- 2024-03-08OAI14 Form 483 citations
What FDA cited at the March 8, 2024 inspection
- 21 CFR 211.84(d)(2)Testing Each Component for Conformity with Specs
- 21 CFR 211.188Prepared for each batch, include complete information
- 21 CFR 211.192Investigations of discrepancies, failures
- 21 CFR 211.165(a)Testing and release for distribution
- 21 CFR 211.100(a)Absence of Written Procedures
- 21 CFR 211.198(a)Complaint Handling Procedure
- 21 CFR 211.42(c)(10)(iv)Environmental Monitoring System
- 21 CFR 211.142(b)Storage under appropriate conditions
- 21 CFR 211.110(a)Written in-process control procedures
- 21 CFR 211.196Distribution Record Requirements
- 21 CFR 211.166(a)Lack of written stability program
- 21 CFR 211.87Retest of approved components/containers/closures
- 21 CFR 211.113(b)Procedures for sterile drug products
- 21 CFR 211.42(c)(10)(v)Cleaning System
The letter
FDA's text of the letter as posted on fda.gov. FDAgent leaves out the address block, the greeting and the copy list, which name the people it was sent to.
The United States Food and Drug Administration (FDA) inspected your facility, located at the above address between February 27, 2024, and March 8, 2024. During the inspection, FDA documented that your company manufactures a human amniotic fluid derived product, Purified Fluid Allograft, as well as a human amniotic membrane derived product, Allograft Matrix, for allogeneic use (collectively, “your products”).1
This letter is to advise you that your products are unapproved new drugs in violation of section 505(a) of the Federal Food, Drug, and Cosmetic Act (FD&C Act), 21 U.S.C. § 355(a). Your products are also unlicensed biological products in violation of section 351(a)(1) of the Public Health Service Act (PHS Act), 42 U.S.C. § 262(a)(1). A biological product for which a biologics license application (BLA) has been approved under section 351(a) of the PHS Act is not required to have an approved application under section 505 of the FD&C Act, 21 U.S.C. § 355; 42 U.S.C. § 262(j). Otherwise, with certain exceptions not applicable here, a new drug may not be introduced or delivered for introduction into interstate commerce without an approved application from FDA in effect, as described in section 505(a) of the FD&C Act. Your introduction or delivery for introduction of your products into interstate commerce, or the causing thereof, is prohibited under sections 301(d) of the FD&C Act, 21 U.S.C. § 331(d).
This warning letter also summarizes significant violations of current good manufacturing practice (CGMP) requirements, including violations of section 501(a)(2)(B) of the FD&C Act, 21 U.S.C. § 351(a)(2)(B), and 21 CFR parts 210 and 211. Because your methods, facilities, or controls for manufacturing, processing, packing, or holding drugs do not conform to CGMP, your products are adulterated within the meaning of section 501(a)(2)(B) of the FD&C Act, 21 U.S.C. § 351(a)(2)(B). Your introduction or delivery for introduction of your products into interstate commerce, or the causing thereof, is a prohibited act under section 301(a) of the FD&C Act, 21 U.S.C. § 331(a).
Unapproved New Drug and Unlicensed Biological Product Violations
Based on information and records reviewed by FDA, your products are intended for use in the diagnosis, cure, mitigation, treatment, or prevention of disease or conditions in humans and/or are intended to affect the structure or function of the body. For example, your products are intended for wound healing, to lower inflammation, or for orthopedic use, as evidenced by the following statements in materials collected during the inspection:
- “Purified Amniotic Fluid Allograft…can be applied directly at the site of injury, pain, or inflammation with a simple injection…Once introduced, it causes a systemic healing cascade in the body, targeting the site of injection.”
- “[Purified Amniotic Fluid Allograft] Applications [include]: o Muscle Tears o Soft Tissue Injuries o Wounds, Acute and Chronic o Osteoarthritis o Joint Injuries / Pain”
- “Allograft [Matrix]…Benefits: o Catalyzes natural healing process to support proper remodeling o Provides specific signaling which lowers the body’s inflammatory response o Mitigates formation of scar tissue”
- “Allograft [Matrix]…Use: o … assists in wound healing, both chronic and acute. o … is used to enhance healing after a primary surgical repair.”
Therefore, your products are drugs as defined in section 201(g)(1) of the FD&C Act, 21 U.S.C. § 321(g)(1), and biological products as defined in section 351(i) of the PHS Act, 42 U.S.C. § 262(i).2
Your Allograft Matrix product is also a human cell, tissue, or cellular or tissue-based product (HCT/P) as defined in 21 CFR 1271.3(d) and is subject to regulation under 21 CFR part 1271, issued under the authority of section 361 of the PHS Act, 42 U.S.C. § 264. HCT/Ps that do not meet all the criteria in 21 CFR 1271.10(a) are not regulated solely under section 361 of the PHS Act and the regulations in 21 CFR part 1271. Unless an exception in 21 CFR 1271.15 applies, such products are regulated as drugs, devices, and/or biological products under the FD&C Act and/or the PHS Act and are subject to additional regulation, including applicable premarket review. Based on a review of relevant materials, your company does not qualify for any exception in 21 CFR 1271.15, and your product fails to meet all criteria in 21 CFR 1271.10(a).
For example, your Allograft Matrix product fails to meet the criterion that the HCT/Ps be “intended for homologous use only,” which means that the “labeling, advertising, or other indications of the manufacturer’s objective intent” demonstrate that the HCT/P is intended to perform “the same basic function or functions in the recipient as in the donor.” 21 CFR 1271.3(c) and 1271.10(a)(2). Your product is not intended solely to perform the same basic function or functions of the HCT/P in the recipient as in the donor (e.g., serving as a barrier for amniotic membrane). Rather, your product is intended for use in catalyzing wound healing and reducing inflammation, which is not a basic function of amniotic membrane in the donor.
Therefore, this HCT/P is not regulated solely under section 361 of the PHS Act, 42 U.S.C. § 264, and the regulations in 21 CFR part 1271.3 See 21 CFR 1271.20. In addition to being regulated under section 361 of the PHS Act and 21 CFR part 1271, your product is regulated as a drug as defined in section 201(g)(1) of the FD&C Act, 21 U.S.C. § 321(g)(1), and a biological product as defined in section 351(i) of the PHS Act, 42 U.S.C. § 262(i), as stated above.
Subject to certain exceptions not applicable here, to lawfully introduce or deliver for introduction into interstate commerce a drug that is a biological product, a valid BLA must be in effect under section 351(a)(1) of the PHS Act, 42 U.S.C. § 262(a)(1). Such licenses are issued only after showing that the product is safe, pure, and potent. Your products are not the subject of an approved BLA.
CGMP Violations
FDA’s inspection of your facility documented evidence of significant CGMP violations. At the conclusion of the inspection, FDA investigators issued a Form FDA-483, List of Inspectional Observations (Form FDA-483).
The CGMP violations applicable to your products include, but are not limited to, the following:
1. Failure to establish and follow appropriate written procedures designed to prevent microbiological contamination of drug products purporting to be sterile, including procedures for validation of all aseptic and sterilization processes, as required by 21 CFR 211.113(b). For example, you failed to adequately validate (i.e., by performing media fill simulations) the aseptic process used to manufacture your Purified Fluid Allograft product. Your product purports to be sterile and is expected to be sterile.
2. Failure to establish written procedures for production and process controls designed to assure that the drug products have the identity, strength, quality, and purity they purport or are represented to possess, as required by 21 CFR 211.100(a). Specifically,
a. The manufacturing process for your Purified Fluid Allograft product has not been adequately validated with respect to identity, strength, quality, and purity. During the inspection, you provided a validation study for your amniotic fluid product, which only tested for sterility. However, this is not sufficient to assure the identity, strength, quality, and purity of your product.
b. The manufacturing process for your Allograft Matrix product has not been validated with respect to identity, strength, quality, and purity.
3. Aseptic processing areas are deficient regarding the system for environmental monitoring to prevent contamination, as required by 21 CFR 211.42(c)(10)(iv). For example, you have not performed non-viable particulate monitoring, personnel monitoring, active or passive viable air sampling, or sampling of critical surfaces for microorganisms in the aseptic processing area in association with each production batch.
4. Aseptic processing areas are deficient regarding the system for cleaning and disinfecting the room and equipment to produce aseptic conditions, as required by 21 CFR 211.42(c)(10)(v). For example, you have not validated your process for cleaning and disinfecting the ISO (b)(4) cleanrooms where your products are manufactured.
5. Written records are not always made of investigations into unexplained discrepancies or the failure of a batch or any of its components to meet specifications whether or not the batch has already been distributed , as required by 21 CFR 211.192. For example, from March 2023 and June 2023, your firm had two sterility failures during finished product testing of your Purified Fluid Allograft product. Although you identified the contaminating organisms, you failed to conduct an investigation to determine the root cause of the failures.
6. Failure to establish and follow a written testing program designed to assess the stability characteristics of drug products and to use results of such stability testing to determine appropriate storage conditions and expiration dates, as required by 21 CFR 211.166(a). Specifically, you assigned a (b)(4) expiration date to your products without supporting data.
Response to the Form FDA-483
We have reviewed your response, dated March 28, 2024, to FDA’s Form FDA-483 in detail. Your response is inadequate because it did not provide sufficient detail for FDA to fully assess the adequacy of your corrective actions. It lacks documentation to demonstrate that you have corrected your violations and fails to include a timeline for completion of all necessary corrective actions. For example, in your response, you note that going forward, any vials still in house from previously opened lots will be discarded. However, your response does not address the rubber stoppers and metal caps, which are used as packaging for your finished amniotic fluid products, that are (b)(4) using a sterilization process that has not been validated. Please be advised that in accordance with 21 CFR 211.113(b), you are required to establish and follow appropriate written procedures designed to prevent microbiological contamination of drug products purporting to be sterile, including procedures for validation of sterilization processes.
Your response also does not address your specific plans for disposition of the inventory of Purified Fluid Allograft or Allograft Matrix at your facility that were manufactured under the conditions described in this letter. Further, your response does not address products that have been distributed and remain in the marketplace.
We acknowledge your commitment to voluntarily suspend production activities until you have made corrections to the observations listed on the Form FDA-483. However, your response does not adequately address your failure to have an IND in effect to study your products addressed in this letter or your lack of an approved BLA to lawfully market your products. As noted above, to lawfully market a drug that is a biological product, a valid biologics license must be in effect. 42 U.S.C. 262(a). Such licenses are issued only after showing that the product is safe, pure, and potent. While in the development stage, such a product may be distributed for clinical use in humans only if the sponsor has an IND in effect for that product, as specified by FDA regulations, that covers such clinical use. 21 U.S.C. 355(i); 42 U.S.C. 262(a)(3); 21 CFR Part 312.
Neither this letter nor the observations noted on the Form FDA-483, which were discussed with you at the conclusion of the inspection are intended to be an all-inclusive list of deficiencies that may exist in connection with your products. You are responsible for investigating and determining the causes of any violations and for preventing their recurrence or the occurrence of other violations. It is your responsibility to ensure full compliance with all applicable requirements in the FD&C Act, PHS Act, and all applicable regulations.
This letter notifies you of our concerns and provides you an opportunity to address them. Failure to adequately address these matters may result in action without further notice, including, without limitation, seizure and/or injunction.
We request that you respond in writing within fifteen (15) working days from your receipt of this letter, outlining the specific steps you have taken or plan to take to address any violations and prevent their recurrence. Include any documentation necessary to show that the matters have been addressed. If you cannot address these matters within fifteen (15) working days, please explain the reason for your delay and the timeframe for completion. If you do not believe your products are in violation of the FD&C Act, PHS Act, or applicable regulations, include your reasoning and any supporting information for our consideration.
Send your electronic response to CBERDCMRecommendations@fda.hhs.gov. If you have questions regarding this letter, contact the Division of Case Management, CBER at (240) 402-9156.
Melissa J. Mendoza Director Office of Compliance and Biologics Quality Center for Biologics Evaluation and Research
_________________________
1 At the time of the inspection, FDA collected evidence that your human amniotic fluid derived product is also referred to as “Evexcell™ Purified Nanoparticle Fluid,” “Purified Amniotic Fluid Allograft,” “Purified Amniotic Fluid,” “Trophic Fluid Allograft,” “Trophic Factor Amniotic Fluid,” “Trophic Factor Purified Fluid Allograft,” and “Cryovial”. FDA also collected evidence that your human amniotic membrane derived product is also referred to as “Allograft Patch,” “Dual Layer Allograft Matrix,” “Dual Layer Allograft Patch,” “(b)(4),” “Patch Matrix,” “Matrix,” “Amniotic Membrane,” “(b)(4) Allograft Patch,” and “Membrane.”
2 We note that your amniotic fluid product, Purified Fluid Allograft, is not an HCT/P because amniotic fluid is a secreted bodily fluid and the definition of HCT/Ps in 21 CFR 1271.3(d)(3) excludes secreted or extracted human products.
3 Because your product fails to meet at least one criterion in 21 CFR 1271.10(a), this letter does not evaluate all other criteria in 21 CFR 1271.10(a).
Ask FDAgent about this warning letter
FDAgent reads the letter, the facility's inspections and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The letter, its text, dates, issuing office and subject come from FDA's warning letter pages.
- FDAgent leaves out the letter's address block, greeting and copy list, which name the people it was sent to, and quotes each numbered violation sentence that cites 21 CFR.
- Facilities come from FDA's compliance action record for the letter's case number, or the one FEI the letter names; inspections and citations come from FDA's Data Dashboard.
- FDAgent is independent of FDA.