Torrent Pharmaceuticals Limited Form 483, April 2017
FDA issued this Form 483 to Torrent Pharmaceuticals Limited in Indrad, India (FEI 3005029956) at the end of a drug inspection that ended April 28, 2017. It lists 4 observations. FDA classified the inspection Voluntary Action Indicated (VAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted May 11, 2017. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Apr 28, 2017
- Drugs · Drug Quality Assurance
- FDA's outcome
- VAI
- Voluntary Action Indicated
- Posted by FDA
- May 11, 2017
- 13 days after the inspection
The 4 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Quality unit procedures not fully followed
The responsibilities and procedures applicable to the quality unit are not fully followed. Specifically, 1. The following deficiencies were noted with the change control program: a. Change control records were not closed in a timely manner. Section 4.0 of SOP CQA-055, Version 4.0, Change Control Management Across all the Locations, effective date 1 DEC 16 states that it is the responsibility of the Head of Quality to complete a 'Periodic review of status of Change control forms for its implementation'. However, according to the Review of Change Control Form for the year 2016, Quarter 4 (Document no. CCFS/INF/16/Q4 approved on 7 FEB 17): Year 2011 2012 2013 2014 Number of Change Control Records that Remain Open 3 65 168 226 No final action has been taken for the changes that have not been implemented, and a final assessment has not been made for those changes that have been implemented. b. A Change Control record was not initiated per SOP CQA-055, Version 4.0, Change Control Management Across all the Locations, effective date 1 December, 2016. The Stability Study Data Sheet, RLT/3476/01, effective date 3 June 2016, for the Dissolution Testing of (b) (4) Tablets (b) (4) (b) (4) mg) was updated to include instructions to ensure that the pH meter is calibrated with buffers at pH (b) (4) and (b) (4) immediately before adjusting the pH of the dissolution media. 2. Batch Record Review by Quality Assurance (QA) was deficient: a. The production batch record for (b) (4) Stage (b) (4) Batch (b) (4) had (b) (4) weights recorded for the standard preparation and (b) (4) weights recorded for the sample preparation. There were (b) (4) printouts for weight, and the two printouts dated 16 OCT 16 were stamped with the word 'DISREGARD'. There was no indication or reference recorded in the batch record stating why these weights were disregarded, why the printouts for the weights dated 18 OCT 16 were accepted, or why the incorrect weights on the previous page were not crossed out. The batch record was reviewed and approved by QA on 20 OCT 16, but this discrepancy was not noted or corrected. b. The production batch record for (b) (4) Finished API batch (b) (4) was missing a page. This discrepancy was noted during production and a deviation was raised. The operator recorded the results on the back of page 44 of the issued batch record. This page was not signed as reviewed or verified by QA. 3. Validation batches that fail to meet specification are subsequently not considered to be validation batches without a justification. For example, Validation Protocol PV/BTD.77.05-4E/15 stated that three consecutive validation batches of (b) (4) were manufactured. However, Validation Report PVR/BTD.77.05-4E/16 only included data from two batches (b) (4) and (b) (4)) and did not evaluate batch (b) (4) which failed to meet specification. 4. Poor documentation practices were observed such as the failure to include details such as product name, product strength, nature of the problem, or summary of the conclusion of the investigation in deviation investigation records and change control records such as CCF/INF/16/0352 and CCF/INF/15/0533.
- 2
Failure to thoroughly investigate discrepancies
There is a failure to thoroughly investigate discrepancies whether or not the batch has been distributed. Specifically, Investigation OOS/IN/A/FP/15/019 dated 10 OCT 15 reported out-of-specification (OOS) results for description for (b) (4) Stage (b) (4) batches (b) (4) and (b) (4). A (b) (4) color was observed. Investigation OOS/IN/A/FP/15/023 dated 7 NOV 15 reported OOS results for description for (b) (4) Stage (b) (4) batches (b) (4), (b) (4) and (b) (4) reported as (b) (4), (b) (4), and (b) (4). The specification for description of (b) (4) Stage (b) (4) was originally (b) (4) to (b) (4). (b) (4) (b) (4) Stage (b) (4) is an intermediate in the production of (b) (4) finished API. (b) (4) Stage (b) (4) material is (b) (4) in nature and is (b) (4). This intermediate product is packaged and sampled (b) (4). The investigation into the observed color changes concluded that the presumptive root cause was exposure to moisture during handling. However, it has not been determined whether: (i) the color change impacts material on the surface, or if the entire batch is affected, and (ii) whether moisture can cause a color change to (b) (4) (b) (4) or (b) (4). Furthermore, there is no color standard for analysts to use for comparison during analysis. Further, per CCF/TRC/15/1649, dated 20 NOV 15, the specification for description was changed to (b) (4) to (b) (4) color powder'.
- 3
Lab equipment used outside calibrated range
Laboratory equipment used for analytical testing was used outside of the calibrated range. Specifically, the HPLCs, GCs and dissolution units located in the API, formulation (finished dosage form) and stability sample quality control testing laboratories were used outside of the calibrated range. For example: a. SOP No.: IPQC-124-02, Annexure VIII, titled "Wavelength Accuracy Check" states that wavelength accuracy for HPLCs should be calibrated at (b) (4) +/- (b) (4) nm. However, according to test method STP No. CTP/1405, Version 00, a wavelength of (b) (4) nm is used to detect degradation products of (b) (4) (b) (4) by HPLC for the (b) (4) and (b) (4) tablets (b) (4) mg, (b) (4) mg - batch (b) (4) using HPLC 1112 located in the Stability Laboratory. b. An unqualified thermocouple (Identification No. TH/1285, Serial No. 32400093WS) was used for the verification of GC column oven temperature (eg. GC-1050 located in formulation laboratory and GC-1882 located at API laboratory). This thermocouple was certified for measurements between 50C to 450C. However, according to SOP IPQC-136-03, titled "Procedure for operation, cleaning and calibration of Gas Chromatograph with head space"; the column oven temperature of the GC should be verified and adjusted with the aid of a calibrated thermocouple at the temperature points of (b) (4) and (b) (4) C. A temperature of (b) (4) C is outside of the certified working range of the thermocouple. c. According to the SOP: CQA-078, Version 03 titled, "Procedure for analysis, reporting and reviewing of analytical data", a (b) (4) working standard solution should be injected after a set of (b) (4) or (b) (4) dissolution samples for quantitative analyses. No (b) (4) working standard solution was used for dissolution tests conducted at the API, formulation (finished dosage form) and stability sample testing laboratory (eg. Dissolution unit: DI/1265 located in formulation laboratory and DI/456 located in stability laboratory).
- 4
No check standard for chromatographic testing
Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, during the establishment of system suitability for chromatographic analyses, the firm did not prepare or use a second working standard solution (Check Standard Solution) to confirm the potency, accuracy and purity of the working standard solutions (Primary Standard Solution). A single standard preparation was used to determine the potency, accuracy and purity of the active pharmaceutical ingredient in the sample solution. For example: for the assay, content uniformity, and degradation product tests of (b) (4) & (b) (4) Tablets ((b) (4) mg/(b) (4) mg) recorded in Analytical Report No. 040000203053, a Check Standard was not used. For the assay, content uniformity, and degradation product tests of (b) (4) & (b) (4) Tablets ((b) (4) mg/(b) (4) mg) recorded in Analytical Report No. 040000203966, a Check Standard was not used.
Other Form 483s FDA posted for this facility
Every Form 483 FDA posts under FEI 3005029956, newest first.
- Form 483, June 2024Inspection ended Jun 12, 2024 · Posted Jun 26, 2024VAI
- Form 483, April 2019Inspection ended Apr 16, 2019 · Posted May 2, 2019OAI
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.