Skip to content

Advocate Lutheran General Hospital Form 483, January 2016

FDA issued this Form 483 to Advocate Lutheran General Hospital in Park Ridge, Illinois (FEI 3005310193) at the end of a device inspection that ended January 11, 2016. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Jan 11, 2016
Devices · Postmarket Assurance: Devices
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Jan 17, 2024
2,928 days after the inspection
Facility
FEI 3005310193
Park Ridge, Illinois

FDA's 7 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Report of Death Within 10 Days

    The user facility did not submit FDA Form 3500A or electronic equivalent to FDA and the device manufacturer within ten working days after becoming aware of information that reasonably suggests that a device has or may have caused or contributed to the death of a patient of the facility.

  2. Reporting All Information Reasonably Known

    The user facility report submitted on FDA Form 3500A did not include all information reasonably known.

  3. Lack of Written MDR Procedures

    Written MDR procedures have not been developed .

  4. Reports and information documentation

    The written MDR procedure does not include documentation and recordkeeping requirements for all Medical Device Reports and information submitted to FDA and device manufacturers .

  5. Adverse events--all info not in file

    MDR event files do not contain or reference all adverse event information in the possession of the reporting entity, including documentation of the deliberations and decision making process used to determine if an event was or was not reportable.

  6. Report of Serious Injury Within 10 Days

    The user facility did not submit FDA Form 3500A or electronic equivalent to the known device manufacturer within 10 working days of becoming aware of information that reasonably suggests that a device has or may have caused or contributed to a serious injury to a patient of the facility.

  7. Failure to use FDA Form 3500A

    An MDR adverse event report was submitted on a form other than FDA Form 3500A (MEDWATCH form) or an approved electronic equivalent.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.