21 CFR 803.32: If I am a user facility, what information must I submit in my individual adverse event reports?
FDA's inspection data lists 1 Form 483 citation of 21 CFR 803.32 at 1 facility, from January 2016 to January 11, 2016.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 1
- At 1 inspection
- Facilities cited
- 1
- Distinct FEIs
- Most recent citation
- Jan 11, 2016
- Advocate Lutheran General Hospital
- Warning letters
- 0
- State a violation of it
The regulation
The current text of 21 CFR 803.32 from the Electronic Code of Federal Regulations.
§ 803.32 If I am a user facility, what information must I submit in my individual adverse event reports?
You must include the following information in your report, if reasonably known to you, as described in § 803.30(b). These types of information correspond generally to the elements of Form FDA 3500A:
(a) Patient information (Form FDA 3500A, Block A). You must submit the following:
(1) Patient name or other identifier;
(2) Patient age at the time of event, or date of birth;
(3) Patient gender; and
(4) Patient weight.
(b) Adverse event or product problem (Form FDA 3500A, Block B). You must submit the following:
(1) Identification of adverse event or product problem;
(2) Outcomes attributed to the adverse event (e.g., death or serious injury). An outcome is considered a serious injury if it is:
(i) A life-threatening injury or illness;
(ii) A disability resulting in permanent impairment of a body function or permanent damage to a body structure; or
(iii) An injury or illness that requires intervention to prevent permanent impairment of a body structure or function;
(3) Date of event;
(4) Date of this report;
(5) Description of event or problem, including a discussion of how the device was involved, nature of the problem, patient followup or required treatment, and any environmental conditions that may have influenced the event;
(6) Description of relevant tests, including dates and laboratory data; and
(7) Description of other relevant history, including preexisting medical conditions.
(c) Device information (Form FDA 3500A, Block D). You must submit the following:
(1) Brand name;
(2) Product Code, if known, and Common Device Name;
(3) Manufacturer name, city, and state;
(4) Model number, catalog number, serial number, lot number, or other identifying number; expiration date; and unique device identifier (UDI) that appears on the device label or on the device package;
(5) Operator of the device (health professional, lay user/patient, other);
(6) Date of device implantation (month, day, year), if applicable;
(7) Date of device explantation (month, day, year), if applicable;
(8) Whether the device is a single-use device that was reprocessed and reused on a patient (Yes, No)?
(9) If the device is a single-use device that was reprocessed and reused on a patient (yes to paragraph (c)(8) of this section), the name and address of the reprocessor;
(10) Whether the device was available for evaluation and whether the device was returned to the manufacturer; if so, the date it was returned to the manufacturer; and
(11) Concomitant medical products and therapy dates. (Do not report products that were used to treat the event.)
(d) Initial reporter information (Form FDA 3500A, Block E). You must submit the following:
(1) Name, address, and telephone number of the reporter who initially provided information to you, or to the manufacturer or distributor;
(2) Whether the initial reporter is a health professional;
(3) Occupation; and
(4) Whether the initial reporter also sent a copy of the report to us, if known.
(e) User facility information (Form FDA 3500A, Block F). You must submit the following:
(1) An indication that this is a user facility report (by marking the user facility box on the form);
(2) Your user facility number;
(3) Your address;
(4) Your contact person;
(5) Your contact person's telephone number;
(6) Date that you became aware of the event (month, day, year);
(7) Type of report (initial or followup); if it is a followup, you must include the report number of the initial report;
(8) Date of your report (month, day, year);
(9) Approximate age of device;
(10) Event problem codes—patient code and device code (refer to the “MedWatch Medical Device Reporting Code Instructions”);
(11) Whether a report was sent to us and the date it was sent (month, day, year);
(12) Location where the event occurred;
(13) Whether the report was sent to the manufacturer and the date it was sent (month, day, year); and
(14) Manufacturer name and address, if available.
21 CFR Part 803: Medical Device Reporting, Subpart CRead on eCFR
Citations by year
Form 483 citations of 21 CFR 803.32 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 1 in 2016
| Year | Form 483 citations |
|---|---|
| 2016 | 1 |
How FDA words this citation
FDA's short descriptions and citation text on these 1 citation, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Reporting All Information Reasonably Known1 citations
Citation text
- “The user facility report submitted on FDA Form 3500A did not include all information reasonably known.”1 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 803.32, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- Advocate Lutheran General Hospital2016-01-11 · FEI 3005310193Reporting All Information Reasonably Known
Often cited at the same inspection
Of the 1 inspection where FDA cited 21 CFR 803.32, the share where it also cited each of these sections.
- 21 CFR 803.11What form should I use to submit reports of individual adverse events and where do I obtain these forms?1 (100%)
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 (100%)
- 21 CFR 803.18What are the requirements for establishing and maintaining MDR files or records that apply to me?1 (100%)
- 21 CFR 803.30If I am a user facility, what reporting requirements apply to me?1 (100%)
Ask FDAgent about 21 CFR 803.32
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
Topics that count this section:Device CAPA and complaints
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 803.32; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.