Skip to content

WEARLINQ INC.

1803 Research Blvd Ste 600, Rockville, Maryland, United States

FEI 3021332567

FDA has one inspection on file for WEARLINQ INC. in Rockville, Maryland (FEI 3021332567). It ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 3 citations from this inspection: 2 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 801, the device labeling rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
1, in 2025
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Jan 21, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Software validation for automated processes

      Software used as part of the quality system has not been validated for its intended use according to an established protocol.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Label to bear a unique device identifier

      The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Jan 21, 2026. It asked for a Form 483.

  • Inspection ending Nov 26, 2025NAI

    1 request. In the inspection history above.

    1. Jan 21, 2026FDA-FOIA-2026-670Form 483· 1 of 5 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 3 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3021332567 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3021332567. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.