Immersivetouch Inc
910 W Van Buren St Ste 715, Chicago, Illinois, United States
FEI 3014833795
FDA has 2 inspections on file for Immersivetouch Inc in Chicago, Illinois (FEI 3014833795), from 2020 to 2023. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.30(i).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 2, 2020–2023
- 0 OAI · 0 VAI · 2 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Jan 29, 2024
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design output - review and approval
Design output was not reviewed and approved before release.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Design input - review and approval
Design input requirements were not reviewed and approved by designated individual(s).
Design validation - software validation not performed
Validation of device software was not performed.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Jan 29, 2024. Each one asked for a Form 483.
Inspection ending Sep 14, 2023NAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Jan 29, 2024FDA-FOIA-2024-976Form 483Closed
- Oct 30, 2023FDA-FOIA-2023-9534Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 2 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.22Quality audit1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3014833795 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3014833795. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.