Vero Biotech Inc
387 Nerem St Nw Ste 125, Atlanta, Georgia, United States
FEI 3014617112DUNS 872672477
Company: VERO BIOTECH INC. · Linked by exact D-U-N-S match (872672477) in FDA registration data
FDA has one inspection on file for Vero Biotech Inc in Atlanta, Georgia (FEI 3014617112). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 4 citations from this inspection: 3 under 21 CFR Part 211, the manufacturing rules for finished drugs, and 1 under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing and analysis, through Dec 31, 2026. Its registrant, VERO BIOTECH INC. (D-U-N-S 872672477), has registered 1 other site.
Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2022
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Computer control of master formula records
Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel.
Control procedures to monitor and validate performance
Control procedures are not established which validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.
Regulations FDA cited here
The 4 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 211.68Automatic, mechanical, and electronic equipment1 of 1
- 21 CFR 211.110Sampling and testing of in-process materials and drug products1 of 1
- 21 CFR 211.192Production record review1 of 1
- 21 CFR 820.30Design controls1 of 1
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing, analysis
- Registration runs through
- Dec 31, 2026
- Registrant
- VERO BIOTECH INC.D-U-N-S 872672477
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Other sites with the same registrant
FDA’s registration file names VERO BIOTECH INC. (D-U-N-S 872672477) as the registrant for this site and the one other below. That exact D-U-N-S number is the only link. A registrant is the firm that filed the registration, which is not always the owner.
- VERO BIOTECH INC.Austell, Georgia · FEI 3027057342Registered; no FDA inspection on file
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. This site’s records are filed under FEI 3014617112 in FDA’s data; other sites under its registrant are listed in summary only. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every inspection and enforcement record on this page is filed under FEI 3014617112. Other sites filed under the same registrant D-U-N-S are listed in summary only. Records filed under another FEI or name are otherwise not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.