HealthDent'l LLC
10052 Bode St Ste E, Plainfield, Illinois, United States
FEI 3013411964
FDA has 2 inspections on file for HealthDent'l LLC in Plainfield, Illinois (FEI 3013411964), from 2018 to 2026. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 6 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 2, 2018–2026
- 0 OAI · 0 VAI · 2 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Jul 7, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Label to bear a unique device identifier
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
- ISO 13485:2016 Clause 7.4.1
Procedures for purchasing process not documented
Procedures to ensure that purchased product conforms to specified purchasing information have not been documented.
Devices subject to device identification GUDID data submission requirements.
The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
- ISO 13485:2016 Clause 7.3.9
Requirements for design and development changes not met before implementation
Design and development changes were not reviewed, verified, validated, and approved before implementation.
- ISO 13485:2016 Clause 7.5.11
Procedures for preserving the conformity of product not documented
Procedures were not documented for preserving the conformity of product to requirements during processing, storage, handling, and distribution.
Show 1 more citation
- ISO 13485:2016 Clause 8.2.4
Internal audit procedures not documented
Procedures to describe the responsibilities and requirements for planning and conducting internal audits and recording and reporting audit results were not documented.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - documentation
The results of design validation, including method(s), the date and the individual(s) performing validation, were not documented in the design history file.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Jul 7, 2026. Each one asked for a Form 483.
Inspection ending Apr 9, 2026NAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 4, 2026. FDAgent holds no copy that FDA posted.
- Jul 7, 2026FDA-FOIA-2026-6457Form 483· 1 of 4 sitesRecords released
- Jun 12, 2026FDA-FOIA-2026-5772Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 3 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 801.20Label to bear a unique device identifier1 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 830.300Devices subject to device identification data submission requirements1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3013411964 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3013411964. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.