Anhui Greenland Biotech Co., Ltd.
No. 5 Kaiyuan Road, Mohekou Industrial Zone, Bengbu, China
FEI 3011252807
FDA has one inspection on file for Anhui Greenland Biotech Co., Ltd. in Bengbu, China (FEI 3011252807). It ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 6 citations from this inspection: 5 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 1, in 2019
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Sampling methods - Lack of or inadequate procedures
Procedures to ensure sampling methods are adequate for their intended use have not been adequately established.
DHR - UDI not included
The device history record does not include, or refer to the location of any unique device identifier (UDI) or universal product code (UPC).
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective evaluation of events that may be subject to medical device reporting requirements.
Design input - Lack of or inadequate procedures
Procedures for design input have not been established.
Show 1 more citation
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.184Device history record1 of 1
- 21 CFR 820.198Complaint files1 of 1
- 21 CFR 820.250Statistical techniques1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3011252807 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3011252807. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.