Mettler-Toledo, LLC
6005 Benjamin Rd, Tampa, Florida, United States
FEI 3006654939
FDA has one inspection on file for Mettler-Toledo, LLC in Tampa, Florida (FEI 3006654939). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 9 citations from this inspection. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2015
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Certification not based on adequate test/testing program
Certification was not based upon a testing program in accordance with good manufacturing practices.
Failure to submit
You did not submit an annual report for products requiring one.
Exposure indication - two independent means
For a cabinet x-ray system, lack of two independent means to indicate when and only when x-rays are being generated.
Failure to submit, distinct marking
You did not submit a required product report prior to the introduction of the product into commerce.
Multiple interlocks on doors
A door of a cabinet x-ray system does not have at least two safety interlocks.
Show 4 more citations
Results of tests
You have not established and maintained records of the results of tests for electronic product radiation safety, including the control of unnecessary, secondary or leakage electronic product radiation.
ID label lacks name and address
An identification label failed to provide the name and address of the manufacturer.
Door interlock dependent upon moving part other than door
A door safety interlock which disconnects the energy supply circuit for the high-voltage generator of a cabinet x-ray system is dependent upon a moving part other than the door.
Manufacture date abbreviated or unapproved date coding
The date of manufacture provided on the identification label is not stated in the prescribed manner.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 1002.10Product reports1 of 1
- 21 CFR 1002.13Annual reports1 of 1
- 21 CFR 1002.30Records to be maintained by manufacturers1 of 1
- 21 CFR 1010.2Certification1 of 1
- 21 CFR 1010.3Identification1 of 1
- 21 CFR 1020.40Cabinet x-ray systems1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3006654939 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3006654939. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.