FDA has 3 inspections on file for Jiangsu Suyun Medical Materials Co., Ltd. in Lianyungang, China (FEI 3003717263), from 2014 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 4 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Dec 9, 2014. No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2014–2026
- 1 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 3
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Jun 18, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.6(b)
Adjustment to measuring equipment was not performed or recorded
Adjustments or readjustments of measuring equipment necessary to ensure valid results were not recorded.
- ISO 13485:2016 Clause 7.3.9
Procedures to control design and development changes not documented
Procedures to control design and development changes were not adequately documented.
- ISO 13485:2016 Clause 7.3.6
Design and development verification not performed
Design and development verification was not adequately performed in accordance with planned and documented arrangements to ensure that design and development outputs met design and development input requirements.
- ISO 13485:2016 Clause 8.4
Procedures for analysis of data not documented
Procedures were not adequately documented to determine, collect, and analyze appropriate data to demonstrate the suitability, adequacy, and effectiveness of the quality management system.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Lack of or inadequate procedures for storage.
Procedures for the control of storage areas and stock rooms have not been adequately established.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Jun 18, 2026. It asked for a Form 483.
Inspection ending Apr 10, 2026VAI
1 request. In the inspection history above.
- Jun 18, 2026FDA-FOIA-2026-5924Form 483· 1 of 2 sitesOpen
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 3 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.150Storage1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003717263 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003717263. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.