Skip to content

APC Medical, Ltd.

68 Tewin Rd, Welwyn Garden City, United Kingdom

FEI 3002806386

FDA has one inspection on file for APC Medical, Ltd. in Welwyn Garden City, United Kingdom (FEI 3002806386). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 4 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.

FDA’s Import Alert 89-04 Red List names this firm at this address. FDA printed no FEI on the entry. No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
1, in 2011
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
89-04Red List
Matched on FDA’s firm name and street address

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    2. Service reports/MDRs/complaints

      Service reports that represent MDR reportable events were not automatically considered complaints and processed in accordance with the requirements of 21 CFR 820.198.

    3. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.

    4. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

Regulations FDA cited here

The 4 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Import alerts

FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.

  • 89-04Red ListDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsImport alert pageAlert on fda.gov

    Date published 2009-09-16

    FDA’s entry
    APC Medical, Ltd. · 68 Tewin Rd, Welwyn Garden City, Hertfordshire UNITED KINGDOM
    FDA’s inspection record for FEI 3002806386
    APC Medical, Ltd. · 68 Tewin Rd, Welwyn Garden City, United Kingdom

    Matched on exact firm name, street address and country. FDA printed no FEI on this entry.

    13 products

    • 73 - - - --Anesthesiology
    • 74 - - - --Cardiovascular
    • 76 - - - --Dental
    Show 10 more products
    • 77 - - - --Ear,Nose And Throat
    • 78 - - - --Gastroenterological & Urological
    • 79 - - - --General & Plastic Surgery
    • 80 - - - --General Hospital/Personal Use
    • 84 - - - --Neurological
    • 85 - - - --Obstetrical & Gynecological
    • 86 - - - --Ophthalmic
    • 87 - - - --Orthopedic
    • 89 - - - --Physical Medicine
    • 90 - - - --Radiology

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002806386 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002806386. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.