Skip to content

J.R. Rand Corp.

100 E Jefryn Blvd Ste S, Deer Park, New York, United States

FEI 2418518

FDA has one inspection on file for J.R. Rand Corp. in Deer Park, New York (FEI 2418518). It ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 7 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
1, in 2009
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality policy and objectives

      The quality policy and objectives were not implemented.

    2. Quality Audit/Reaudit - conducted

      Quality audits were not conducted to verify that the quality system is effective in fulfilling your quality system objectives.

    3. Management review dates

      Dates of management reviews are not documented.

    4. Test and inspection data

      Service reports do not include applicable test and/or inspection data.

    5. Device identification, control numbers

      The device history record does not include any device identifications and control numbers used.

    Show 2 more citations
    1. Servicing individual

      Service reports do not include the name of service person.

    2. Management representative

      No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.

Regulations FDA cited here

The 4 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2418518 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2418518. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.