South Bend Medical Foundation, Inc.
3355 Douglas Rd, South Bend, Indiana, United States
FEI 1870308
FDA has 8 inspections on file for South Bend Medical Foundation, Inc. in South Bend, Indiana (FEI 1870308), from 2009 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published one citation from this inspection, under 21 CFR Part 606, the manufacturing rules for blood and blood components. It is listed below. Two of the four FOIA requests for this site have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 8, 2009–2025
- 0 OAI · 8 VAI · 0 NAI
- Inspections with 483 observations shown
- 7 of 8
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 4 requests, latest Dec 31, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection and processing of blood and blood components for allogeneic transfusion were not always followed.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
SOPs for investigations, records
Written standard operating procedures for all steps in recordkeeping related to current good manufacturing practice and other applicable requirements and standards were not followed.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Cleaned, sanitized per established schedules
Equipment used for manufacturing HCT/Ps was not cleaned according to established schedules.
Adverse Reaction- Reports of Investigations
Written reports of investigations of adverse reactions, including conclusions and follow up, are not prepared and maintained.
Equipment observed, standardized, calibrated
Equipment used in the collection of blood and blood components is not observed on a regularly scheduled basis as prescribed in the SOP Manual.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Thorough investigations
Failure to perform a thorough investigation of an unexplained discrepancy.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Thorough investigations
Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.
Person performing, test results, interpretation
Records fail to identify the person performing the work, include dates of the various entries, and be as detailed as necessary so as to provide a complete history of the work performed.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection, processing, compatibility testing, storage, and distribution of blood and blood components for homologous transfusion, autologous transfusion, and further manufacturing purposes are not always maintained and followed.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed.
Adverse Reaction- Reports of Investigations
Written reports of investigations of adverse reactions, including conclusions and follow up, are not prepared and maintained.
Show 2 more citations
Equipment observed, standardized, calibrated
Equipment used in the processing of blood and blood components is not standardized and calibrated on a regularly scheduled basis as prescribed in the SOP Manual.
Record review prior to release
All records pertinent to a lot or unit were not reviewed before the release or distribution of a lot or unit of final product.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection, processing, storage, and distribution of blood and blood components for homologous transfusion and further manufacturing purposes are not always maintained and followed.
Adverse Reaction - Investigations
A thorough investigation of each reported adverse reaction was not made.
Equipment calibration frequency
Equipment used in the collection of blood and blood components is not observed, standardized and calibrated with at least the frequency required.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Donor suitability procedures not followed
Failure to follow standard procedures and methods for determining the suitability of a donor as a source of blood.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection of blood and blood components for homologous transfusion are not always maintained and followed.
Thorough investigations
Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy and a failure of a lot or unit to meet any of its specifications.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Dec 31, 2025. Each one asked for a Form 483.
Inspection ending Aug 1, 2025VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Mar 30, 2026. FDAgent holds no copy that FDA posted.
- Dec 31, 2025FDA-FOIA-2025-14418Form 483Records released
- Sep 24, 2025FDA-FOIA-2025-11223Form 483· 1 of 5 sitesRecords released
Inspection ending Sep 1, 2023VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 16, 2023FDA-FOIA-2023-9057Form 483Closed
Inspection ending Aug 9, 2019VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Jan 30, 2024FDA-FOIA-2024-1007Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 7 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 8 inspections, with how many of them cited each.
- 21 CFR 606.100Standard operating procedures6 of 7
- 21 CFR 606.60Equipment3 of 7
- 21 CFR 606.170Adverse reaction file3 of 7
- 21 CFR 211.22Responsibilities of quality control unit1 of 7
- 21 CFR 606.160Records1 of 7
- 21 CFR 640.31 of 7
- 21 CFR 1271.200Equipment1 of 7
Other FEIs recorded as South Bend Medical Foundation, Inc.
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
South Bend Medical Foundation, Inc.FEI 3000204209South Bend, Indiana, United States
Last inspected 2026-01-07VAI
South Bend Medical Foundation, Inc.FEI 1872515Goshen, Indiana, United States
Last inspected 2024-02-01NAI
South Bend Medical Foundation, Inc.FEI 3007701926Marion, Indiana, United States
Last inspected 2014-06-24NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1870308 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1870308. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.