LEVO AG
Angliker Strasse 20, Wohlen Ag, Switzerland
FEI 1000268398
FDA has one inspection on file for LEVO AG in Wohlen Ag, Switzerland (FEI 1000268398). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 5 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, 1 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 830, the unique device identification rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
1 import refusal is filed under this FEI, on . FDA’s Import Alert 89-04 Red List names this firm at this address. FDA printed no FEI on the entry. No court action is filed under this FEI.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 1, in 2025
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Sep 10, 2025
- Warning letters
- · CDRH ↗
- Import refusals
- 1 refusal
- Import alerts
- 89-04Red List
- Matched on FDA’s firm name and street address
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality System Procedures
Quality system procedures and instructions have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained and implemented.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Devices subject to device identification GUDID data submission requirements.
The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
Personnel
Personnel do not have the necessary background, training and experience to perform their jobs.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Sep 10, 2025. It asked for a Form 483.
Inspection ending Apr 9, 2025OAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Feb 26, 2026. FDAgent holds no copy that FDA posted.
- Sep 10, 2025FDA-FOIA-2025-10795Form 483· 1 of 2 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 820.20Management responsibility1 of 1
- 21 CFR 820.25Personnel1 of 1
- 21 CFR 820.198Complaint files1 of 1
- 21 CFR 830.300Devices subject to device identification data submission requirements1 of 1
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1000268398, one row per refused product line.
1 refusal on 2019-08-15 · charged NOT LISTED on 1, FRNMFGREG on 1 · no FDA sample analysis
FDA firm profile- 2019-08-15
WHEELCHAIR, STANDUP89IPL
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Northern Border Imports785-5032171-1/21/2
Import alerts
FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.
- 89-04Red ListDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsImport alert pageAlert on fda.gov
Date published 2025-08-14
- FDA’s entry
- LEVO AG · Angliker Strasse 20, Wohlen Ag, Aargau SWITZERLAND
- FDA’s inspection record for FEI 1000268398
- LEVO AG · Angliker Strasse 20, Wohlen Ag, Switzerland
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
1 product
- 89 I - - PLWheelchair, Standup
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000268398 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000268398. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.