Skip to content

21 CFR 212.71: What actions must I take if a batch of PET drug product does not conform to specifications?

FDA's inspection data lists 19 Form 483 citations of 21 CFR 212.71 at 15 facilities, from August 2016 to July 23, 2026.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
19
At 16 inspections
Facilities cited
15
Distinct FEIs
Most recent citation
Jul 23, 2026
Ionetix Corporation
Warning letters
0
State a violation of it

The regulation

The current text of 21 CFR 212.71 from the Electronic Code of Federal Regulations.

§ 212.71 What actions must I take if a batch of PET drug product does not conform to specifications?

(a) Rejection of nonconforming product. You must reject a batch of a PET drug product that does not conform to specifications. You must have and follow procedures to identify and segregate the product to avoid mix-ups. You must have and follow procedures to investigate the cause(s) of the nonconforming product. The investigation must include, but is not limited to, examination of processes, operations, records, complaints, and any other relevant sources of information concerning the nonconforming product.

(b) Investigation. You must document the investigation of a PET drug product that does not meet specifications, including the results of the investigation and what happened to the rejected PET drug product.

(c) Correction of problems. You must take action to correct any identified problems to prevent recurrence of a nonconforming product or other quality problem.

(d) Reprocessing. If appropriate, you may reprocess a batch of a PET drug product that does not conform to specifications. If material that does not meet acceptance criteria is reprocessed, you must follow procedures stated in the product's approved application and the finished product must conform to specifications, except for sterility, before final release.

21 CFR Part 212: Current Good Manufacturing Practice for Positron Emission Tomography Drugs, Subpart HRead on eCFR

Citations by year

Form 483 citations of 21 CFR 212.71 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 5 in 2025

Form 483 citations of 21 CFR 212.71 by inspection year
YearForm 483 citations
20161
20183
20191
20201
20233
20255
20265

How FDA words this citation

FDA's short descriptions and citation text on these 19 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Correction of problems9 citations
  • Documentation of Non-Conforming Product Investigation4 citations
  • Non-Conforming Product Investigation Incomplete3 citations
  • Investigation of nonconforming product1 citations
  • Procedures to Investigate Cause of Non-Conforming Product1 citations
Show 1 more
  • Rejection of nonconforming product1 citations

Citation text

  • “You did not take appropriate action to correct any identified problems to prevent recurrence of a nonconforming product or other quality problem.”7 citations
  • “You did not document the results of the investigation for a PET drug product that did not meet specifications.”2 citations
  • “You did not take action to correct any identified problems to prevent recurrence of a nonconforming product or other quality problem.”2 citations
  • “The investigation of the cause of the nonconforming batch of a PET drug product did not include examination of the other relevant sources of information concerning the nonconforming product.”1 citations
Show 2 more
  • “The investigation of the cause of the nonconforming batch of a PET drug product did not include examination of the processes, operations and other relevant sources of information concerning the nonconforming product.”1 citations
  • “The investigation of the cause of the nonconforming batch of a PET drug product did not include examination of the processes, operations, records and other relevant sources of information concerning the nonconforming product.”1 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 212.71, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

Show 4 more
  • Kettering Medical Center2025-05-09 · FEI 3008961887Documentation of Non-Conforming Product Investigation
  • Ionetix Corporation2025-05-01 · FEI 3017953815Correction of problems
  • PETNET Solutions, Inc2023-09-08 · FEI 3007134238Non-Conforming Product Investigation Incomplete
  • SOFIE Co2023-08-11 · FEI 3008415560Documentation of Non-Conforming Product Investigation

Often cited at the same inspection

Of the 16 inspections where FDA cited 21 CFR 212.71, the share where it also cited each of these sections.

  • 21 CFR 212.20What activities must I perform to ensure drug quality?7 (44%)
  • 21 CFR 212.30What requirements must my facilities and equipment meet?7 (44%)
  • 21 CFR 212.50What production and process controls must I have?4 (25%)
  • 21 CFR 212.60What requirements apply to the laboratories where I test components, in-process materials, and finished PET drug products?4 (25%)
  • 21 CFR 212.100What do I do if I receive a complaint about a PET drug product produced at my facility?3 (19%)
  • 21 CFR 212.70What controls and acceptance criteria must I have for my finished PET drug products?3 (19%)

Ask FDAgent about 21 CFR 212.71

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 212.71; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.