Olympus Medical Systems Corporation warning letter, December 12, 2022

FDA's Center for Devices and Radiological Health issued this warning letter to Olympus Medical Systems Corporation of Hachioji-shi, Tokyo on December 12, 2022 and posted it on January 10, 2023. It states one violation, citing 21 CFR 803.50(a)(2).

Source: FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Read the full letter on fda.gov. FDAgent is independent of FDA.

Issued
Dec 12, 2022
By CDRH
Posted by FDA
Jan 10, 2023
29 days after issue
Subject
Medical Device Reporting/​Misbranded
As FDA lists it
FDA reference
MARCS-CMS 649726
FDA's case number

Violations FDA states in this letter

Each numbered violation sentence that cites 21 CFR, quoted exactly from the letter. The full letter below explains each one. A check means FDA's citation data also lists that section for the facility's September 9, 2022 inspection.

  1. 1
    “Failure to submit a report to FDA no later than 30 calendar days after the day that your firm received or otherwise became aware of information, from any source, that reasonably suggests that a device that your firm markets has malfunctioned and this device or a similar device that it markets would be likely to cause or contribute to a death or serious injury, if the malfunction were to recur, as required by 21 CFR 803.50(a)(2).”
    21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?993 Form 483 citations · 33 warning letters✓ Cited at the September 9, 2022 inspection

The facility

FDA's compliance action record for this letter (case 649726) names one facility. Each facility's FDA record on FDAgent has its full inspection history.

Inspections

  • 2025-11-21OAI3 Form 483 citations
  • 2022-09-09OAI1 Form 483 citation
  • 2015-04-24OAI2 Form 483 citations

What FDA cited at the September 9, 2022 inspection

The letter

FDA's text of the letter as posted on fda.gov. FDAgent leaves out the address block, the greeting and the copy list, which name the people it was sent to.

During an inspection of your firm located in Hachioji-shi, Tokyo, Japan, on September 6, 2022, through September 9, 2022, an investigator (or investigators) from the United States Food and Drug Administration (FDA) determined that your firm manufactures gastrointestinal and surgical endoscopes. Under section 201(h) of the Federal Food, Drug, and Cosmetic Act (the Act), 21 U.S.C. § 321(h), these products are devices because they are intended for use in the diagnosis of disease or other conditions or in the cure, mitigation, treatment, or prevention of disease, or to affect the structure or any function of the body.

Our inspection revealed that your firm's gastrointestinal and surgical endoscopes are misbranded under section 502(t)(2) of the Act, 21 U.S.C. § 352(t)(2), in that your firm failed or refused to furnish material or information respecting the device that is required by or under section 519 of the Act, 21 U.S.C. § 360i, and 21 CFR Part 803 - Medical Device Reporting. We received a response from your firm, dated September 30, 2022, to the Form FDA 483 (FDA 483). We address this response below, in relation to each of the noted violation(s). These violation(s) include, but are not limited to, the following:

1. Failure to submit a report to FDA no later than 30 calendar days after the day that your firm received or otherwise became aware of information, from any source, that reasonably suggests that a device that your firm markets has malfunctioned and this device or a similar device that it markets would be likely to cause or contribute to a death or serious injury, if the malfunction were to recur, as required by 21 CFR 803.50(a)(2).

For example, the information in complaint (b)(4) (MDR# 8010047-2021-02641), complaint (b)(4) (MDR# 8010047-2021-07387), complaint (b)(4) (MDR# 8010047-2021-11050), and complaint (b)(4) (MDR# 8010047-2021-09059), reasonably suggests that your firm’s Uretero-Reno Fiberscope and Uretero-Reno Videoscope malfunctioned (e.g., device breakage or rupture) while in use. Your firm failed to submit an MDR for each of the referenced complaints within the required 30-calendar day timeframe. As a result of these complaints, your firm conducted recall Z-2518-2018, for the Uretero-reno fiberscope URF-P6, where it was established that such events could result in potential perforation of a patient’s organs. Per 21 CFR 803.3, a malfunction is reportable if the manufacturer takes or would be required to take an action under sections 518 or 519(g) of the Act as a result of the malfunction of the device or other similar devices. As such, each of the referenced complaints represents an MDR-reportable malfunction, as defined in 21 CFR 803.3.

The adequacy of your firm’s response dated September 30, 2022, cannot be determined at this time as the response notes that your firm is continuing to implement its systemic corrective actions, including optimizing the MDR reporting process and staff training. However, the adequacy of your firm’s response cannot be determined because your firm’s actions are presently ongoing. When your firm’s corrective actions are completed, please provide documentation or evidence of the systemic corrective actions, including a copy of your revised MDR procedure and a retrospective review of complaints for reportability.

U.S. federal agencies may take your compliance history into account when considering the award of contracts.

Please notify this office in writing, within fifteen business days from the date you receive this letter, of the specific steps your firm has taken to correct the noted violation, as well as an explanation of how your firm plans to prevent the violation, or similar violation, from occurring again. Include documentation of the corrections and/or corrective actions (which must address systemic problems) that your firm has taken. If your firm's planned corrections and/or corrective actions will occur over time, please include a timetable for implementation of those activities. If corrections and/or corrective actions cannot be completed within fifteen business days, state the reason for the delay and the time within which these activities will be completed. Your firm's response should be comprehensive and address all violation included in this Warning Letter. If you believe that your products are not in violation of the FD&C Act, include your reasoning and any supporting information for our consideration as part of your response. Please provide a translation of documentation not in English to facilitate our review. We will notify you regarding the adequacy of your firm's response(s) and the need to re-inspect your firm's facility to verify that the appropriate corrections and/or corrective actions have been made.

Your firm's response should be sent to CDRHWarningLetterResponses@fda.hhs.gov, or if e-mail cannot be used:

Food and Drug Administration - CDRH/ORP/DRP2 CDRH Regulatory Inspections and Audit Team, WO66-1434 10903 New Hampshire Ave. Silver Spring, MD 20993

Refer to CMS case # 478315

If you have any questions about the contents of this letter, please contact: Mark Antonino at Mark.Antonino@fda.hhs.gov

Finally, you should know that this letter is not intended to be an all-inclusive list of the violation at your firm's facility. It is your firm's responsibility to ensure compliance with applicable laws and regulations administered by FDA.

Sincerely Yours, /S/

Courtney H. Lias, Ph.D. Office Director OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Ask FDAgent about this warning letter

FDAgent reads the letter, the facility's inspections and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The letter, its text, dates, issuing office and subject come from FDA's warning letter pages.
  • FDAgent leaves out the letter's address block, greeting and copy list, which name the people it was sent to, and quotes each numbered violation sentence that cites 21 CFR.
  • Facilities come from FDA's compliance action record for the letter's case number, or the one FEI the letter names; inspections and citations come from FDA's Data Dashboard.
  • FDAgent is independent of FDA.