Jiangsu Shenli Medical Production Co., Ltd. warning letter, March 18, 2024
FDA's Center for Devices and Radiological Health issued this warning letter to Jiangsu Shenli Medical Production Co., Ltd. of Changzhou Shi, Jiangsu Sheng on March 18, 2024 and posted it on March 19, 2024. FDA lists its subject as CGMP/QSR/Medical Devices/PMA/Adulterated/Misbranded.
Source: FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Read the full letter on fda.gov. FDAgent is independent of FDA.
- Issued
- Mar 18, 2024
- By CDRH
- Posted by FDA
- Mar 19, 2024
- 1 day after issue
- Subject
- CGMP/QSR/Medical Devices/PMA/Adulterated/Misbranded
- As FDA lists it
- FDA reference
- MARCS-CMS 677753
- FDA's case number
Violations FDA states in this letter
This letter states no violation sentence that cites a 21 CFR section FDAgent can quote. The full letter is below.
The facility
FDA's compliance action record for this letter (case 677753) names one facility. Each facility's FDA record on FDAgent has its full inspection history.
Inspections
- 2024-03-29OAINo Form 483 citations listed
- 2015-05-07VAI1 Form 483 citation
- 2012-11-15NAINo Form 483 citations listed
What FDA cited at the May 7, 2015 inspection
- 21 CFR 820.70(e)Contamination control, Lack of or inadequate procedures
The letter
FDA's text of the letter as posted on fda.gov. FDAgent leaves out the address block, the greeting and the copy list, which name the people it was sent to.
The United States Food and Drug Administration (FDA) has learned that your firm is marketing Jiangsu Shenli Medical Production Co. Ltd. piston syringes of the sizes and configurations listed in Table 1 below in the United States without marketing clearance or approval, in violation of the Federal Food, Drug, and Cosmetic Act (the Act).
Item Description
Item Description
SYR 20ML L/L YELLOW
SYR 3ML L/L PURPLE
SYR 5ML L/L PURPLE
SYR 10ML L/L PURPLE
SYR 20ML L/L BLUE SALINE
SYR 10ML L/L YELLOW CONT
SYR 10ML L/L BLUE LIDO
SYR 10ML L/L BLUE SALINE
SYR 10ML L/L RED LIDO
SYR 10ML L/L RED HEPARIN
SYR 5ML L/L YELLOW NITRO
SYR 10ML L/L YELLOW LIDO
SYR 3ML L/L RED
SYR 10ML L/L YELLOW HEP/SALINE
SYR 3ML L/L GREEN
SYR 10ML L/L RED CONT
SYR 3ML L/L BLUE
SYR 10ML L/L GREEN CONT
SYR 5ML L/L RED
SYR 20ML L/L WHITE SALINE
SYR 5ML L/L GREEN
SYR 5ML L/L PURPLE NITRO
SYR 5ML L/L BLUE
SYR 20ML L/L YELLOW LIDO
SYR 5ML L/L WHITE
SYR 5ML L/L RED NITRO
SYR 5ML L/L YELLOW
SYR 5ML L/L BLUE HEPARIN
SYR 10ML L/L RED
SYR 10ML L/S BLUE
SYR 10ML L/L GREEN
SYR 10ML L/L WHITE LIDO
SYR 10ML L/L BLUE
SYR 10ML L/S RED
SYR 10ML L/L WHITE
SYR CNTRL 10ML L/L YEL 1
SYR 10ML L/L YELLOW
SYR CNTRL 10ML L/IL YELLOW LIDO
SYR 20ML L/L RED
SYR CNTRL 10 ML L/L RED
SYR 20ML L/L GREEN
SYR CNTRL 10ML L/L GRN
SYR 20ML L/L BLUE
SYR CNTRL 10ML L/L BLUE
SYR 20ML L/L WHITE
SYR 10ML L/L CONTROL
Table 1. List of Jiangsu Shenli Medical Production Co. Ltd. piston syringe configurations implicated in this Warning Letter.
Under section 201(h) of the Act, 21 U.S.C. § 321(h), these products are devices because they are intended for use in the diagnosis of disease or other conditions or in the cure, mitigation, treatment, or prevention of disease, or to affect the structure or any function of the body.
FDA has determined, based on information provided by your firm and information available on the Agency's public databases, that the Jiangsu Shenli Medical Production Co. Ltd. piston syringes configurations listed in Table 1 are adulterated under section 501(f)(1)(B) of the Act, 21 U.S.C. § 351(0(1)(B), because your firm does not have an approved application for premarket approval (PMA) in effect pursuant to section 515(a) of the Act, 21 U.S.C. § 360e(a), or an approved application for an investigational device exemption (IDE) under section 520(g) of the Act, 21 U.S.C. § 360j(g). The devices are also misbranded under section 502(o) the Act, 21 U.S.C. § 352(o), because your firm introduced or delivered for introduction into interstate commerce for commercial distribution these devices with (1) major changes or modifications to the intended use, and/or (2) a change or modification in the device that could significantly affect the safety or effectiveness of the device without submitting a premarket notification to FDA as required by section 510(k) of the Act, 21 U.S.C. § 360(k), and 21 CFR 807.81(a)(3)(i)-(ii).
Specifically, the Jiangsu Shenli Medical Production Co. Ltd. Piston Syringe (5cc Luer Lock) was cleared under K103830 with the following indications for use: The intended use of the Jiangsu Shenli Medical Production Co., Ltd. 5cc luer lock piston syringe is to inject fluids into or withdraw fluids from the body. Based on FDA's evaluation of your firm's activities, there is evidence that your firm is engaged in the distribution of syringes for use within the United States with substantially different technological characteristics, namely sizes other than 5 mL, luer slip instead of luer lock tips, syringes intended for specific drugs, different colors (i.e., colorants), and control syringes, each of which constitute a significant change or modification in design that could significantly alter the safety or effectiveness of the device. For further explanation on the need for a new premarket notification, or "510(k)," for changes to the intended use or design affecting safety and effectiveness, it is recommended you consult the guidance document "Deciding When to Submit a 510(k) for a Change to an Existing Device — Guidance for Industry and Food and Drug Administration Staff (fda.gov)." In the guidance at Figure 3 — Flowchart B: Technology, Engineering, and Performance Changes, FDA explains why changes in size, color, specific drug indications and connector type all constitute a major change or modification to the device's intended use, for which your firm lacks clearance or approval.
As discussed above the products listed in Table 1 are misbranded because you made one or more change(s) or modification(s) to each device that could significantly affect the safety or effectiveness of the device without submitting a 510(k) to FDA. Specifically, changing the volume of the syringe, switching from a luer lock to a luer slip syringe tip, and changing from a traditional syringe to a control syringe could result in the risk of patient harm such as inaccurate dosing, a leaking device, a higher risk of inappropriate needle detachment, and incorrect device handling. Furthermore, the addition of color additives to your syringe could result in adverse health effects such as allergic reactions, skin irritation or inflammation, pain, fever, red blood cell damage that induces organ stress, and toxicity that leads to loss of organ function or failure. Labeling the syringe for a specific drug can also cause harm because differences in drug viscosities impact the function of the syringe and can result in patient harm such as inaccurate dosing or incorrect drug usage.
Your firm's failure to submit 510(k)s for these devices has prevented FDA from evaluating the risks posed by these substantially different technological characteristics or from determining that there is reasonable assurance of the safety and effectiveness of these modified devices such that they may be legally marketed syringes. Further, FDA has observed evidence of postmarket safety signals indicating product quality issues that have the potential to cause serious patient harm. Specifically, FDA is aware of multiple customer complaints regarding quality issues (e.g. cracks, breakage during use) for syringes in surgical kits supplied by Jiangsu Shenli Medical Production Co. Ltd.
For a device requiring premarket approval, the notification required by section 510(k) of the Act, 21 U.S.C. § 360(k), is deemed satisfied when a PMA is pending before the agency. 21 CFR 807.81(b). The kind of information that your firm needs to submit in order to obtain approval or clearance for the device is described on the Internet at http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/HowtoMarketYourDevice/default.htm. The FDA will evaluate the information that your firm submits and decide whether the product may be legally marketed.
FDA requests that Jiangsu Shenli Medical Production Factory cease any activities that result in the misbranding or adulteration of all the Jiangsu Shenli Medical Production Co. Ltd. piston syringe configurations listed in Table 1, such as the commercial distribution of the device sizes, configurations, and uses discussed above, including selling to initial importers, distributors, or other third parties for distribution and use within the United States or otherwise representing these devices to have the requisite clearance or authorization for marketing in the United States.
Given the serious nature of the violations of the Act and growing evidence of potential harm, all the Jiangsu Shenli Medical Production Co. Ltd. piston syringes configurations listed in Table 1 are subject to refusal of admission under section 801(a) of the Act, 21 U.S.C. § 381(a), in that they appear to be adulterated. As a result, FDA is taking steps to refuse entry of these devices into the United States, until these violations are addressed.
Please notify this office in writing within fifteen business days from the date you receive this letter of the specific steps your firm has taken to address the noted violations, as well as an explanation of how your firm plans to prevent these violations, or similar violations, from occurring again. Include documentation of the corrections and/or corrective actions (which must address systemic problems) your firm has taken. If your firm's planned corrections and/or corrective actions will occur over time, please include a timetable for implementation of those activities. If corrections and/or corrective actions cannot be completed within fifteen business days, state the reason for the delay and the time within which these activities will be completed. Your firm's response should be comprehensive and address any violations included in this Warning Letter. If you believe that your products are not in violation of the Act, include your reasoning and any supporting information for our consideration as part of your response.
Your firm's response should be sent by email to CDRHWamingLetterResponses@fda.hhs.gov or by mail to:
Food and Drug Administration Center for Devices and Radiological Health Office of Regulatory Programs Division of Regulatory Programs 2: Establishment Support Regulatory Inspections and Audits Team White Oak Building 66 10903 New Hampshire Ave. Silver Spring, MD 20993
Refer to the identification number CMS # 677753 when replying. We remind you that only written communication is considered as official. If you have any questions about the contents of this letter, please contact: Shruti Misty at Shruti.Mistry@fda.hhs.gov.
Finally, you should know that this letter is not intended to be an all-inclusive list of the violations at your firm. It is your firm's responsibility to ensure compliance with the applicable laws and regulations administered by FDA.
Sincerely yours, /S/
Kellie B. Kelm, Ph.D. Acting Director OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Official correspondent: Cai Xiaotian at (b)(4)
US Agent: Charles Mack IRC USA 2950 E Lindrick Dr Chandler, AZ, US 85249 Email: (b)(4) or (b)(4)
Ask FDAgent about this warning letter
FDAgent reads the letter, the facility's inspections and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The letter, its text, dates, issuing office and subject come from FDA's warning letter pages.
- FDAgent leaves out the letter's address block, greeting and copy list, which name the people it was sent to, and quotes each numbered violation sentence that cites 21 CFR.
- Facilities come from FDA's compliance action record for the letter's case number, or the one FEI the letter names; inspections and citations come from FDA's Data Dashboard.
- FDAgent is independent of FDA.