Surgery Pharmacy Services, Inc. Form 483, December 2019

FDA issued this Form 483 to Surgery Pharmacy Services, Inc. in Chattanooga, Tennessee (FEI 3010704905) at the end of a drug inspection that ended December 6, 2019. It lists 10 observations. FDA classified the inspection Official Action Indicated (OAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jan 29, 2021. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Dec 6, 2019
Drugs · Drug Quality Assurance
FDA's outcome
OAI
Official Action Indicated
Posted by FDA
Jan 29, 2021
420 days after the inspection
Facility
FEI 3010704905
Chattanooga, Tennessee

The 10 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Microbial contamination, inadequate response

    Actionable microbial contamination was present in production without adequate product evaluation and remedial action. Specifically, on 5/27/2019, response to Bacillus microbial contamination [illegible] the extent of impact nor identification and notification of all potentially impacted patients.

  2. 2

    Infrequent glove disinfection/change

    Personnel failed to disinfect or change gloves frequently enough. Specifically, the technician was observed during preparation of TPN products on 11/18/2019. For example, she moved her hands out of the ISO5 hood and then back into the hood without sanitizing, tying out the trash bag in the large trash bin located in the ISO8 anteroom, and resuming aseptic production activities in the ISO5 hood with hands with [illegible]. She did not change her sterile gloves after contact with the trash bin. On 11/19/2019, she was also observed using a step-stool, which she uses to reach within the ISO5 hood, and did not change her sterile gloves before resuming aseptic production activities.

  3. 3

    Exposed skin in critical area

    Personnel engaged in aseptic processing were observed with exposed skin in critical areas. On 11/18/2019 and 11/19/2019, the technician was observed leaning significantly into the ISO5 certified hood with her forehead exposed while leaning during aseptic production activities.

  4. 4

    Materials not adequately disinfected

    Equipment, materials, and/or supplies are not adequately disinfected prior to entering aseptic processing areas. Specifically, materials are not adequately disinfected to stricter classifications. Large drug components, such as [illegible], are loaded and then moved into temporary storage baskets within the pharmacy area prior to being placed into the ISO5 hood. For production, materials are placed on ISO5 worksurfaces prior to disinfection.

  5. 5

    Inadequate media fill simulations

    Media fills were not performed that closely simulate appropriate, worst-case activities and conditions. Specifically, the media fills consist of preparing and [illegible] vials intended to represent finished product, though activities such as the manual production and [illegible] for intracameral injections are not represented.

  6. 6

    Non-sterile disinfectants used

    Disinfecting agents used in the ISO5 area are [illegible]; [illegible] is non-sterile and is routinely used in supporting aseptic processing areas.

  7. 7

    Unverified pressure differentials

    Pressure differentials between areas with different air classifications were not monitored prior to or during sterile drug production. Specifically, there is no verification of the pressure differentials before or during sterile drug production. Pressure differentials are measured with manometers which are not visible from within the cleanroom. The values are documented but not recorded [illegible].

  8. 8

    Released drug failed assay testing

    Your firm released drug product in which the strength differs from, or its purity or quality falls below, that which it purports or is represented to possess. Specifically, epinephrine solutions have failed assay testing for epinephrine since 02/2019 but they were still produced and dispensed without resolving the failures. Additionally, testing data lacks supporting documentation with the date of production to verify the solutions.

  9. 9

    Damaged, hard-to-clean hood surface

    The ISO-classified equipment have difficult-to-clean surfaces. Specifically, the ISO5 hood has a worksurface composed of laminate that is stained on the left side and has been chipped on both front corner edges. This hood is used for sterile drug production and was the only hood in use through 12/2/2019.

  10. 10

    Certification not representative of operations

    ISO-5 classified areas were not certified or verified under operational conditions representative of dynamic conditions. The smoke studies performed only include evaluation of the hood when not fully loaded and do not include evaluation of a fully-loaded hood, such as when equipment takes up a significant amount of space and potentially interrupts the unidirectional airflow hood design.

After this inspection

FDA's compliance actions for this FEI dated after this inspection and before its next one. FDAgent links a letter when FDA's case number ties it to this facility.

  • Warning LetterOct 9, 2020CDER · 308 days after this inspection ended

Other Form 483s FDA posted for this facility

Every Form 483 FDA posts under FEI 3010704905, newest first.

All inspections at this facility →

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.