Skip to content

Hospira Inc. Form 483, October 2012

FDA issued this Form 483 to Hospira Inc. in Mcpherson, Kansas (FEI 1925262) at the end of a drug inspection that ended October 24, 2012. It lists 2 observations. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Apr 21, 2021. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Oct 24, 2012
Drugs · Drug Quality Assurance
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Apr 21, 2021
3,101 days after the inspection
Facility
FEI 1925262
Mcpherson, Kansas

The 2 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Inadequate aseptic gowning and traffic

    Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established. Specifically, the design of the personnel entryways, personnel and material traffic flows and gowning practices, particularly with regard to dedicated plant uniforms, captive plant shoes and shoe covers, is deficient. The current [illegible] promotes the propagation of microbial contamination. The following was observed during the inspection 09 OCT - 24 OCT 2012 and has been the current practice for approximately one (1) year. This is applicable to all drug products, over approximately two hundred drug product combinations of strengths and dosage configurations. Aseptic filling personnel were observed to frequent common public areas such as administrative offices, the warehouse, and the cafeteria without being required to change out of dedicated plant shoes which then were worn back into aseptic fill rooms under their aseptic gowning. Employees and visitors wearing shoes which are worn outside the facility must don shoe covers in manufacturing corridors and hallways. However, employees in dedicated plant shoes were seen coming from the warehouses, and in the cafeteria, commingling with personnel wearing outside shoes then returning to the plant manufacturing areas for work. There is no distinction in these practices other than a [illegible] sanitization step required [illegible] of all employees in dedicated shoes. This sanitization is negated as soon as employees return to common hallways and public areas of the plant. This is a failure to mitigate the ingress of dirt, debris and microorganisms. For June, July and August of 2011 and 2012 there was an increase over action limit trending guidelines used by the firm for the presence of microorganisms in environmental monitoring data which represents a seasonal increase in the recovery of vegetative organisms.

  2. 2

    Investigations not extended facility-wide

    Investigations of an unexplained discrepancy did not extend to other drug products that may have been associated with the specific failure or discrepancy. Specifically, 1. Since March 2012 your firm has discovered 2 lots [illegible] of Fentanyl Citrate for injection which failed HPLC stability time point analysis for total impurities. Subsequent investigation of the failures [illegible] were due to extractables from the rubber plunger. In addition, under study your firm tested [illegible] lots of Fentanyl Citrate from retain samples and determined one additional failed for total impurities and 2 others were at the USP limit. There are [illegible] lots of Fentanyl Citrate currently distributed and unexpired; your investigation was not expanded to comprehensively determine the number of lots which fail USP specifications. 2. There were 2 incidents within the past 2 months of city water leaking from sprinkler heads located in your aseptic areas; these sprinkler heads and piping had been replaced by your contractor during a recent shutdown. • Incident PRID: 95301 occurred 17 Aug 2012 near lyophilizer [illegible] (room 335, 335B) during the unloading of [illegible] drug product. The root cause of the leak was attributed to the tapering of the threading between the piping and the reducer; the pipes did not fit each other correctly. • Incident PRID: 95564 occurred 20 Aug 2012 in the SPA corridor near the entrance door to the [illegible] filling suite during filling of Diphenhydramine. The root cause was attributed to the pipe fitting seal on the sprinkler system plumbing. Although in each case the event was contained, and drug product was not affected, neither investigation extended to other wet sprinkler heads throughout the facility to adequately evaluate if they had been properly installed, or could be leaking.

Other Form 483s FDA posted for this facility

Every Form 483 FDA posts under FEI 1925262, newest first.

All inspections at this facility →

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.