Skip to content

Guidemia Technologies LLC Form 483, July 2026

FDA issued this Form 483 to Guidemia Technologies LLC in Buena Park, California (FEI 3013634985) at the end of a device inspection that ended July 6, 2026. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Nov 20, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Jul 6, 2026
Devices · Postmarket Assurance: Devices
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Nov 20, 2026
137 days after the inspection
Facility
FEI 3013634985
Buena Park, California

FDA's 4 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. ISO 13485:2016 Clause 8.2.2

    Complaint handling records not maintained

    Complaint handling records were not adequately maintained.

  2. GUDID Information required for unique device identification.

    The correct unique device identifier (UDI) information regarding device identifier portion of the UDI assigned to the version or model was not provided to the FDA via the Global Unique Device Identification Database (GUDID).

  3. ISO 13485:2016 Clause 8.5.2

    Corrective action records not maintained

    Records of corrective actions, including the results of any investigations and actions taken, were not adequately maintained.

  4. ISO 13485:2016 Clause 6.2(b)

    Personnel training not provided

    Training has not been adequately provided, or other actions needed taken, to achieve or maintain the necessary personnel competence.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.