Dr. Reddy's Laboratories Ltd. Form 483, June 2019
FDA issued this Form 483 to Dr. Reddy's Laboratories Ltd. in Visakhapatnam, India (FEI 3006549835) at the end of a device inspection that ended June 21, 2019. It lists 2 observations. FDA classified the inspection Voluntary Action Indicated (VAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Jul 9, 2019. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Jun 21, 2019
- Devices · Compliance: Devices
- FDA's outcome
- VAI
- Voluntary Action Indicated
- Posted by FDA
- Jul 9, 2019
- 18 days after the inspection
The 2 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Validation acceptance criteria missing
Acceptance criteria were not established prior to the performance of validation activities. 4 A Ps . : (b) (4) . Specifically, is not clear how your firm determined the acceptance criteria for the Functionality Testing (Break loose force &, Glide force) conducted as part of finished product release specification for IB Injection®® mg@nL. For example, your firm explained that the worst-case condition was considered for the applied injection @@ “based on the container size as part of the criteria to test the anples. Your data was only obtained 4% the” “samples from stability study for 24 months for’ (Batches: ed as reference data for Break loose force of @ Newton and Glide force of Newton. ~ . PR) (b) . However, your firm established the test limit for Break loose force as @ Newton and for Glide force as Newton as the finished product release specification test limits. This was not supported by scientific data that demonstrated the impact to the end user or to the finished product functionality.
- 2
Design not translated into specs
The device design was not correctly translated into production specifications. Specifically, your Acceptance Quality Limit [AQL] does not take into account the component critical dimensions specification during receipt of components for the manufacture of Injection ©@ ho @nL such as®® and stoppers.
Other Form 483s FDA posted for this facility
Every Form 483 FDA posts under FEI 3006549835, newest first.
- Form 483, October 2021Inspection ended Oct 29, 2021 · Posted Dec 6, 2021VAI
- Form 483, August 2019Inspection ended Aug 20, 2019 · Posted Aug 29, 2019VAI
- Form 483, October 2018Inspection ended Oct 30, 2018 · Posted Nov 14, 2018VAI
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.